Two-point calibration
(using a set of two calibration foils):
a) Recommended for measurements on rough surfaces
b) Recommended for precise measurements on smooth surface if the
thickness to be expected lies between that of the two calibration foils
Storing Calibration Values
If the meter is calibrated for a particular purpose, the calibration values
will be stored in memory until changed.
The calibration procedure should be restarted from the beginning if:
- An incorrect reading has been taken
- An incorrect command has been entered
- The meter has been switched off
Calibration Example
Calibration is the required for accurate measurements. The more closely
the calibration sample matches the product sample, the more accurate
the calibration, and therefore the reading will be. If for instance, a product
is to be measured on a steel cylinder, quality ST37 (mild steel), diameter
6mm, the calibration of the uncoated sample must take place on a steel
cylinder of similar quality with the same diameter. The calibration sample
must correspond to the product sample in the following ways:
- Curvature radius
- Substrate material properties
- Substrate thickness
- Size of measuring area
- The point at which the calibration is made on the calibration sample
must always be identical with the point of measurement on the product
itself, especially in the case of corners and edges of small parts