1)  Turn on the meter by pressing the Power Button. To turn the meter 

off, hold the Power Button for over 2 seconds.

2)  Plug in the BNC Cable into the BNC Socket, and the Mini Cable 

Plug into the Vibration Sensor Socket.

3)  If the surface material of measuring article is not ferrous, hold the 

Vibration Sensor and touch the Sensor to the surface of the mea-
suring article. If the surface material of measuring article is ferrous, 
connect Magnetic Base to the Vibration sensor. Place the Vibration 
sensor and Magnetic Base on the surface of measuring article. Do 
not touch the Sensor Cable when taking measurements.

Changing Unit

To change the measuring unit, press and hold the Unit Button. The display 
will flash the units m/s

2

, g, mm/s, cm/s, ft/s

2

, inch/s. Release the Unit 

Button to select the displayed unit. When measuring with an Accelera-
tion Unit, the screen will display “ACC”. When measuring with an Velocity 
Unit, the screen will display “VEL”. The selected unit will be saved as the 
default. 

Function selection 

1)  The meter has 3 measuring functions to choose from: RMS, Peak, 

and Max Hold.

2)  When the meter is on, press and hold the Function Button. The 

display will flash “RMS”, “Max Hold”, and “Peak”. Release the 
Function Button to choose the displayed function.

3)  RMS measures and Root Means Square (RMS) value. Max Hold 

measures and updates the max. peak value. Peak measures the 
peak vibration value. For Acceleration and Velocity measurement, 
always select RMS measurement. The selected function will be 
saved as the default.

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Содержание SD-8205

Страница 1: ...20 20 20 100 Only if you REALLY need them Model Vibration Meter Datalogger SD 8205 reedinstruments www com Instruction Manual www GlobalTestSupply com Find Quality Products Online at sales GlobalTest...

Страница 2: ...t 7 Max hold reset procedures 7 Data Hold 7 Data Record 7 8 Backlight ON OFF 8 Datalogger 8 Auto Datalogger 8 9 Manual Datalogger 9 Checking the Time 9 Check Sampling Time Information 9 SD Card Data S...

Страница 3: ...No extra software needed just plug the SD card into the computer and it downloads directly to EXCEL SD card capacity 1 GB to 16 GB Easy to read LCD with backlight Auto power off or manual power off D...

Страница 4: ...case and batteries Optional accessories 4G SD Card SD 4GD External Probe SD 8205PROBE USB cable USB 01 Power adapter AP 9VA 110V Data Acquisition Software SW U801 WIN Tripod BS 6 Visit us at www reedi...

Страница 5: ...er ESC Backlight Button 3 Hold Function Next Button 4 REC Enter Unit Button 5 SET Down Time Check Button 6 Logger Up Sampling Time Check Zero Max Button 7 BNC input socket 8 SD card slot 9 USB RS232 O...

Страница 6: ...h the Sensor Cable when taking measurements Changing Unit To change the measuring unit press and hold the Unit Button The display will flash the units m s2 g mm s cm s ft s2 inch s Release the Unit Bu...

Страница 7: ...s and hold the Logger Button for at least 5 seconds to clear the existing Max hold value Data Hold While taking a measurement press the Hold Button once and the mea sured value will hold on the screen...

Страница 8: ...nsert the optional SD card into the SD card socket 2 Before you use an SD card with this meter for the first time it is recommend to format the SD card See Advanced Settings section of this manual for...

Страница 9: ...d Press the REC Button and the REC symbol will appear on the LCD screen Press the Logger Button and the REC symbol will flash once and one Beep will sound at the same time the measuring data along the...

Страница 10: ...1 can hold 99 XLS files When that number has been reached the meter will then will generate a new folder with a subsequent number for example VBA02 Max number of VBA folders is 10 Data Transfer from S...

Страница 11: ...ing function press the Esc Button to exit and to return to normal functioning Set Clock Time 1 When the lower display shows dAtE press the Enter Button Use the Up and Down Buttons to adjust the Year A...

Страница 12: ...yES Beeper Sound is ON or no Beeper Sound is OFF 2 After selecting yES or no press the Enter Button to save the setting as the default To exit Advanced Settings press the Esc Button Set sampling time...

Страница 13: ...5 mm terminal The connector output is a 16 digit data stream which can be utilized to the user s specific application An USB RS232 lead with the following connection will be required to link the inst...

Страница 14: ...rt Word System Reset If an error occurs press the Reset Button with a Pin If the Reset Button does not fix the error please send in the unit to our Calibration Lab for repair E mail info reedinstrumen...

Страница 15: ...act REED Instruments at info reedinstruments com Notes __________________________________________ ________________________________________________ ________________________________________________ ____...

Страница 16: ..._____________________________ ________________________________________________ ________________________________________________ ________________________________________________ _______________________...

Отзывы: