12
4. Press the and arrows to select either "ON" or "OFF".
5. Press the
button to confirm the value.
6.
Press the
ESC
button to exit the setup menu and resume normal operation.
Setting Average Measurement Time
The processing measurement time can be set from 8 to 120 seconds.
Note:
An increase of radiation will automatically and proportionally reduce
the average measurement time. When the time is set to 8 seconds and the
radiation strength is more than 5μSv/h, the response time will be adjusted
to 2 seconds. The factory default time is set to 30 seconds. Follow the
steps below to set the average measurement time.
1. Press the
SETUP
button to enter the set up mode.
2. Press the and arrows to highlight the icon.
3. Press the
button to enter the average measurement time
setup screen.
4. Make sure to start with the leftmost digit.
5. To toggle between digits press the
button to move to the next
digit or press the
ESC
button to return to the previous digit.
6. When ready, press the and arrows to adjust the flashing digit.
7. Press the
button to confirm your selection and proceed to the
next digit.
8. When the last digit has been selected and confirmed press the
ESC
button to exit the setup menu and resume normal operation.
Turning Bluetooth Data Transmission ON/OFF
1. Press the
SETUP
button to enter the set up mode.
2. Press the and arrows to highlight
.
3. Press the
button to enter the Bluetooth data transmission
activation screen.
4. Press the and arrows to select either "ON" or "OFF".
5. Press the
button to confirm the value.
6.
Press the
ESC
button to exit the setup menu and resume normal operation.
7. The icon will appear on the main screen confirming.