9
Auto Power OFF
To preserve battery life, the instrument is programmed to turn itself OFF
after 5 minutes of inactivity. Prior to the instrument turning OFF, the LCD
display will blink for approximately 30 seconds. Press the
button to
delay the auto power OFF function and resume normal operation.
Resistance Measurement Mode
1. When instrument is turned on and passes calibration, "OL Ω" appears
on the LCD display as shown below confirming that you can now
measure the resistance.
2. Pull the trigger to open the clamp and clamp onto the calibration
loop being measured.
3. The LCD will display the resistance value.
4. If the instrument beeps and the LCD display flashes, the resistance
measurement is exceeding the critical set alarm. Refer to
Setting the
Critical Alarms
for details.
Current Measurement Mode
1. After the instrument has powered up and passed calibration, press
the
A
button to enter the current measurement mode.
2. The LCD display will now indicate AC 0.00mA as shown below.
continued...