Fault Diagnostics
Symptom
Problem
Action
"OL" appears on
display when
measuring
Measured value
is greater than the
maximum range
Stop measuring
"-OL" appears
on display when
measuring
Measured value is
less than the minimum
range
Stop measuring
Err appears when
booting
Exceeding the
minimum or maximum
operating ambient
temperature
Place the thermometer
in a 0°C to 50°C(32°F
to 122°F) environment
and resume operation
after 30 minutes
Low battery
indicator flashes
Low battery
Replace the battery
Laser fails to work/
weak laser
Low battery
Replace the battery
Inaccurate
measurement
Wrong emissivity
setting, measured
distance is too far,
measured target
diameter is less than
20mm, etc.
Verify field of view and
the distance to spot
of target
Product Care
To keep your instrument in good working order we recommend the following:
• Store your product in a clean, dry place.
• Change the battery as needed.
• If your instrument isn't being used for a period of one month or longer
please remove the battery.
• Clean your product and accessories with biodegradable cleaner. Do not
spray the cleaner directly on the instrument. Use on external parts only.