Страница 1: ...Infrared Thermometer Instruction Manual R2330 www com information itm com 1 800 561 8187 ...

Страница 2: ...erage and Differential Functions 10 Turning Laser On Off 10 Setting the High Temperature Alarm 11 Setting the Low Temperature Alarm 11 Adjusting Emissivity 12 Switching the Unit of Measure F C 12 Audible Alarm Setting 13 Trigger Lock Setup 13 Turning High and Low Alarms ON OFF 13 14 Battery Replacement 15 Applications 15 Appendix Emissivity Chart 16 18 Fault Diagnostics 18 Product Care 18 Product ...

Страница 3: ...REED distributor or authorized Service Center Please note an additional fee for this service will apply Safety Never attempt to repair or modify your instrument Dismantling your product other than for the purpose of replacing batteries may cause damage that will not be covered under the manufacturer s warranty Servicing should only be provided by an authorized service center Use extreme caution wh...

Страница 4: ...od mount for continuous monitoring Low battery indication and auto shut off Included Carrying Case Battery Specifications Temperature Range 26 to 2282 F 32 to 1250 C Accuracy Basic 1 8 of reading or 3 6 F 1 8 C Detailed 0 C 1 8 C or 1 8 of reading whichever is greater 0 C 1 8 C 0 1 C C 32 F 3 6 F or 1 8 of reading whichever is greater 32 F 3 6 F 0 1 F F Resolution 0 1 F 0 1 C Optical Resolution D ...

Страница 5: ... Internal Memory Yes Tripod Mountable Yes Laser Class Class II Low Battery Indicator Yes Power Supply 9V Battery Battery Life Approx 7 hours with continuous use backlight on Product Certifications CE RoHS Operating Temperature 32 to 122 F 0 to 50 C Storage Temperature 4 to 140 F 20 to 60 C Operating Humidity Range 10 to 90 Maximum Operating Altitude 6561 2000m Maximum Storage Altitude 39370 12000m...

Страница 6: ...00 62 1500 R2320 Professional InfraredThermometer www REEDINSTRUMENTS com 1 Dual Laser Pointer Beams 2 IR Sensor 3 Trigger 4 Battery Cover 5 Tripod Mount 6 LCD Display 7 Mode Button Down Arrow 8 Set Button 9 High Low Alarm Button Up Arrow 10 Laser ON OFF Button www com information itm com 1 800 561 8187 ...

Страница 7: ...ator Temperature Measurement Alarm Indicator Low Battery Temperature Measurement Active Measurement Data Hold Unit of Measurement Temperature Measurement Value Temperature Measurement Value from Selected Mode Emissivity Laser On Mode K K www com information itm com 1 800 561 8187 ...

Страница 8: ...8 Distance Spot Size 4 8 8 8 7 www com information itm com 1 800 561 8187 ...

Страница 9: ... measurement time to be set from 1 minute to 4 hours Refer to Trigger Lock Setup for details Note This function is applicable for processes that require regular monitoring of temperature 1 Once the trigger lock function is enabled the icon will appear on the display and the SCAN icon will flash when the trigger is pressed 2 Pull the trigger again to stop measuring and both the and SCAN icons will ...

Страница 10: ...sponding modes are shown in the secondary display area Turning Laser On Off Press the button to turn the laser indicator function ON and OFF When it is turned on the laser indicator will be displayed on the LCD while the red laser beam will indicate the position you are measuring during testing continued www com information itm com 1 800 561 8187 ...

Страница 11: ...he setup after 10 seconds of inactivity or if the trigger is pulled at any time Setting the Low Temperature Alarm Press the SET button twice to enter the low alarm limit setup The meter will automatically default to the previously selected preset Use the UP or DOWN buttons to set the temperature value within selected the preset Press the UP or DOWN buttons to add or subtract by 1 temperature value...

Страница 12: ...emissivity values Press the UP and DOWN buttons to add or subtract 0 01 unit and or hold the UP and DOWN buttons to add or subtract 0 1 The instrument will exit the setup after 10 seconds of inactivity or if the trigger is pulled at any time Switching the Unit of Measure F C Press the SET button four times to enter unit of measure setup Use the UP and DOWN buttons to switch between F and C units T...

Страница 13: ...To adjust press the UP or DOWN buttons Press the a third time to confirm the entered time stamp The instrument will exit the setup after 10 seconds of inactivity or if the trigger is pulled at any time Turning High and Low Alarms ON OFF Press the HI LO button to enable or disable the set alarms To set alarms refer to Setting the High Temperature Alarm or Setting the Low Temperature Alarm When the ...

Страница 14: ...temperature value is normal When the HI LO limit alarm function is turned on and the measured temperature value is within the high and low alarm limit range the green LED lights up and the OK indicator is displayed indicating that the measured temperature value is normal 1 High Low alarms on 2 High alarm on 3 Low alarm on 4 No alarm set 3 4 www com information itm com 1 800 561 8187 ...

Страница 15: ... Verify mechanical bearings motors or electrical circuit breaker boxes equipment Calibration and control of heater and oven temperatures Monitoring materials in processes involving heating and or cooling Quality control monitoring Research and development applications Automotive diagnostics Conduct HVAC R energy audits www com information itm com 1 800 561 8187 ...

Страница 16: ...rneath it Measure the temperature of the tape or painted surface Measured Surfaces Emissivity Metal Aluminum Oxidization 0 2 0 4 A3003 Alloy Oxidization Rough 0 3 0 1 0 3 Brass Burnishing Oxidization 0 3 0 5 Copper Oxidization Electric Terminal Board 0 4 0 8 0 6 Hastelloy Alloy 0 3 0 8 Inconel Oxidization Sand Blasting Electro Burnishing 0 7 0 95 0 3 0 6 0 15 Iron Oxidization Rusting 0 5 0 9 0 5 0...

Страница 17: ...95 Basalt 0 7 Carbon Non Oxidization Graphite Silicone Carbide 0 8 0 9 0 7 0 8 0 9 Ceramics 0 95 Clay 0 95 Concrete 0 95 Cloth 0 9 Glass Convex Smooth Lead Boron 0 76 0 8 0 92 0 94 0 78 0 82 Plates 0 96 Stone Products 0 93 Plaster 0 8 0 95 Ice 0 98 Limestone 0 98 Paper 0 95 Plastics 0 95 Water 0 93 Soil 0 9 0 98 Wood 0 9 0 95 www com information itm com 1 800 561 8187 ...

Страница 18: ... battery Laser fails to work weak laser Low battery Replace the battery Inaccurate measurement Wrong emissivity setting measured distance is too far measured target diameter is less than 20mm etc Verify field of view and the distance to spot of target Product Care To keep your instrument in good working order we recommend the following Store your product in a clean dry place Change the battery as ...

Страница 19: ...e at 1 877 849 2127 or by email at info reedinstruments com to discuss the claim and determine the appropriate steps to process the warranty Product Disposal and Recycling Please follow local laws and regulations when disposing or recycling your instrument Your product contains electronic components and must be disposed of separately from standard waste products Product Support If you have any que...