Millivolt Calibration (CAL)

Connect precision millivolt source with an accuracy of 0.03% to terminals #8 

(-) & #9 (+). Cold Junction or RTD ohms calibration MUST be performed after 

millivolt calibration.

Thermocouple Cold Junction Calibration (CJC)

This procedure must be performed AFTER an accurate mV calibration.

1. Exit Factory Service Operations (continually press P until “End”), and return 

to Normal Display Mode.

2. Connect a thermocouple probe of known accuracy to the TLA (Types T, E, J, 

K, and N only). Select the probe type used in Configure Module 1.

3. Connect a reference temperature probe to the measuring end of the TLA 

thermocouple probe. The two probes should be shielded from air movement 

and allowed sufficient time to equalize in temperature. (As an alternative, the 

TLA thermocouple probe may be placed in a calibration bath of known 

temperature.)

4. Compare TLA display with reference temperature probe (or calibration bath). 

If the displayed TLA temperature does not equal the reference probe 

temperature, calculate the CJ error as follows:

CJ Error = reference probe temperature - displayed TLA temperature

5. Enter Factory Service Operations Module (9-FS).

RTD Ohms Calibration (RTD)

This procedure must be performed AFTER an accurate mV calibration. 

Connect one leg of precision resistance (accuracy of 0.1 ohm) to terminals #9 

and #10 together, and the other leg to #8.

12

6.0 C

ALIbRATION

When re-calibration is required (generally every two years), this procedure 

should be performed by qualified technicians using appropriate equipment.  

Equipment source accuracy of 0.03% or better is required.

The procedure consists of: applying accurate mV signals, setting the 

thermocouple cold junction temperature, and applying precision resistance, 

among others. Allow a 30 minute warm-up period before starting this procedure. 

Do not use thermocouple wire for the millivolt or RTD ohms calibration.

This procedure may be aborted by disconnecting power to the TLA before 

exiting the configuration mode. The existing calibration settings remain in 

affect.

Note: After completing any of the calibration sequences, the TLA defaults the 

input sensor type to thermocouple type “J” (tc-J). Be sure to set input sensor 

for proper type.

Note: The TLA must be restored to normal display mode before any data is 

stored.

Not required if only using TC input. This procedure can only be performed 

AFTER an accurate mV calibration.

yes/no

RTD resistance calibration

rtd

Not required if only using RTD input. This procedure can only be performed 

AFTER an accurate mV calibration.

yes/no

Thermocouple cold junction 

temperature calibration

CJC

Calibration required for both RTD and TC input. If this procedure is performed, 

the cold junction temp or RTD ohms calibration procedures in turn must be 

completed.

Calibrate instrument.

yes/no

48

Millivolt calibration

Enter function code

CAL

CodE

DESCRIPTION/ COMMENTS

RANGE

PARAMETER

DISPLAY

Factory Service Operations - Calibration (9-FS)

Wait 10 seconds, press P.

Apply 14.0 mV

Wait 10 seconds, press P.

Wait 10 seconds, press P.

Apply 56.0 mV

Apply 42.0 mV

StP5

StP4

StP2

Wait 10 seconds, press P.

Apply 28.0 mV

StP3

Apply 0.0 mV

StP1

DESCRIPTION/ COMMENTS

ACTION

DISPLAY

Wait 10 seconds, press P.

Cold Junction 

Temperature

CJC

DESCRIPTION/ COMMENTS

PARAMETER

DISPLAY

Note: If the initial value for CJC is not within 

the range of 15°C to 40°C, enter 25.0° for 

CJC and repeat the Cold Junction Calibration 

procedure.

Observe the indicated cold junction 

temperature. Add the calculated CJ Error to 

the displayed value. Enter the sum as the new 

value for CJC. Exit 9-FS and repeat step 4.

Wait 10 seconds, press P.

rtd2

Wait 10 seconds, press P.

rtd1

DESCRIPTION/ COMMENTS

ACTION

DISPLAY

Connect 0.0 ohm 

(jumper wire)

Connect 277.0 ohm

Содержание TLA

Страница 1: ...to activate according to a variety of actions Absolute HI or LO Deviation HI or LO and Band IN or OUT with adjustable hysteresis A standby feature suppresses the alarm during power up until the process stabilizes outside the alarm region The unit is constructed of a lightweight high impact plastic case with a tinted front panel The front panel meets NEMA 4X IP65 specifications when properly instal...

Страница 2: ...nel latch Panel latch can be installed for vertical or horizontal instrument stacking One piece tinted plastic bezel Bezel assembly with circuit boards can be removed from the case to change the output board without removing the case from the panel or disconnecting wiring Unit meets NEMA 4X IP65 requirements for indoor use when properly installed Flame resistant Installation Category II Pollution ...

Страница 3: ...ded This is especially useful in applications where cable runs are long and portable two way radios are used in close proximity or if the installation is near a commercial radio transmitter Also Signal or Control cables within an enclosure should be routed as far away as possible from contactors control relays transformers and other noisy components 4 Long cable runs are more susceptible to EMI pi...

Страница 4: ... ventilation to ensure that the maximum operating temperature range is not exceeded Caution Disconnect power to the unit and to the output control circuits to eliminate the potential shock hazard when removing the bezel assembly Unit Removal Procedure To remove a unit from the panel first loosen the panel latch screws Insert flat blade screwdrivers between the latch and the case on either side of ...

Страница 5: ...ging applications two or more thermocouple probes may be connected to the TLA always use the same type Paralleling a single thermocouple to more than one TLA is not recommended Generally the red wire from the thermocouple is negative and connected to the TLA s common RTD When connecting the RTD be certain that the connections are clean and tight RTD sensors have a higher degree of accuracy and sta...

Страница 6: ...ode In this mode the process temperature is displayed in the main top display and the limit setpoint is displayed in the secondary bottom display When changes to the parameter configurations are needed the P button is pressed and the TLA will enter into the Parameter Mode PARAMETER CONFIGURATION BASIC STARTUP For basic start up it is important to verify or change Input Parameter Module 1 IN parame...

Страница 7: ...egree resolution If changed be sure to check all parameters FLtr Digital Input Filtering and Display Update Select the relative degree of input signal filtering and display update rate The filter is an adaptive digital filter that discriminates between measurement noise and actual process changes Therefore the influence on step response time is minimal If the signal is varying too greatly due to m...

Страница 8: ...n low acting with positive alarm value d LO deviation high acting with negative value d HI or band inside acting b IN is selected for the alarm action the indicator is OFF when the alarm output is ON See the Alarms section for complete details of each action If changed check alarm values The alarm values are either absolute values or relative to the limit setpoint value deviation and band alarms A...

Страница 9: ...r this flag perform calibration procedure as noted in the Calibration section Alternatively stepping through one of the calibration procedures clears the error flag but does NOT validate the calibration accuracy in any manner USER PARAMETER VALUE CHART TLA Number _____________ MNEMONIC PARAMETER USER SETTING Limit Setpoint Alarm 2 Value Password Alarm 1 Value Input Parameters MNEMONIC PARAMETER US...

Страница 10: ... ON OFF Energized High Trip Action Low Trip Action 5 0 Alarms Optional The alarm action figures describe the status of the alarm output and the front panel indicator for various over under temperature conditions The alarm output wave form is shown with the output in the automatic reset mode Select the alarm action with care in some configurations the front panel indicator LED might be OFF while th...

Страница 11: ...e The TLA display should equal the calibrated probe temperature Tolerance is 1ºC 4 Calibrate the cold junction temperature if out of tolerance RTD Ohms Reading Check 1 Connect RTD simulator with an accuracy of 0 1 ohm or better capable of operating with less than 150 µA to terminals 8 9 10 2 Configure Input Parameters Module 1 for linear ohms rLin input under tYPE 3 Compare the TLA read out with t...

Страница 12: ...rm up period before starting this procedure Do not use thermocouple wire for the millivolt or RTD ohms calibration This procedure may be aborted by disconnecting power to the TLA before exiting the configuration mode The existing calibration settings remain in affect Note After completing any of the calibration sequences the TLA defaults the input sensor type to thermocouple type J tc J Be sure to...

Страница 13: ...ISPLAY 1 Input is below range of TLA 2 Temperature below range of input probe 3 Defective or incorrect transmitter or probe 4 Excessive low temperature for probe 5 Loss of setup parameters OLOL OR ULUL IN LOWER DISPLAY 1 Signal input exceeds allowable range by 5 1 Check remote signal source 1 Check input parameters 2 Change to input sensor with a lower temperature range 3 Replace transmitter or pr...

Страница 14: ...t Board RBDLA210 AL1 LIMIT E B 0 7 8 4 MD1797 A2 A1 OUT EX P OUTPUT BOARD POWER CPU BOARD SUPPLY BOARD OUTPUT BOARD CONNECTOR BEZEL DISPLAY BOARD ASSEMBLY BEZEL BOARD LATCH DISPLAY BOARD CONNECTOR 9 0 Terminal Configurations 50 60 HZ 8VA 14 12 85 250 VAC AC 1 A2 N O 13 COMM COMM A1 N O LIMIT RELAY ALARM RELAYS 4 N O 5 AC 11 2 3 6 REMOTE RESET 9 TC 10 RTD 8 7 COMM 50 60 HZ 9VA AC 24V 10 DC 18 36V 7...

Страница 15: ...15 This page intentionally left blank ...

Страница 16: ...h Buyer its employees or sub contractors are or may be to any extent liable including without limitation penalties imposed by the Consumer Product Safety Act P L 92 573 and liability imposed upon any person pursuant to the Magnuson Moss Warranty Act P L 93 637 as now in effect or as amended hereafter No warranties expressed or implied are created with respect to The Company s products except those...

Отзывы: