REBOTEC / BA No.: 02.07.147/2017-01 

 GB

 

- 14 - 

Do not use an abrasive scouring agent or cleaner
(read  the  product  instructions!)  that  contain
aggressive  solvents  (acids,  benzene,  thinners,
alcohol).

Do not use stiff cleaning objects (e.g. brushes), as
these  will  damage  the  plastic  surfaces  or  cause
deposits to build up on the plastic parts and other
materials.

Attention!  Before  any  cleaning,  press  the 
emergency stop switch.

 

Protection 
class 

Cleaning instructions 

IPX5 

Washable with water and brush but 
without running water or immersion 
bath. 

IPX6 

Washable with water and brush. 
Water can be used with pressure. 
The system must not be exposed 
directly to the high-pressure cleaner 
jet. Max 20°C (68° F). 

Keep the product clean, and clean it on a regular 
basis.  Hygiene  must  be  monitored  carefully, 
especially in sensitive areas. 
In order  to  ensure  the  protection  class  (IPX5  & 
IPX6),  all  electrical  components  must  be 
installed  with  each  other  and  none  of  the 
connections (C) must be open. 

14. Maintenance and upkeep

The cause  of accidents or  damage may be  traced 
back  to  a  late  detection  of  wear  and/or  improper 
use, as well as irregular maintenance and upkeep. 

For  safety  reasons  and  for  preventing 
accidents,  the  chair  must  undergo  a 

complete  maintenance

 

at  least

 

once  a 

year

 (refer to the Medical Products Act and 

local requirements).  

14.1 Maintaining the chair 

Regular safety-technical checks are necessary:

The product for visible damage (cracks,
fractures, loose parts, instability).

All screwed connections for tightness (especially
the castor wheels and frame).

Function of the armrests and latching pins.

Function of the footrests (replace defective foot
plates).

Rolling ability, check function of the wheels
(replace defective wheels).

Function of the castor locking brakes (replace
defective brakes).

Function of the headrest.

Function of the chair inclination.

Check accessories for inadequacies.

In the event of any inadequacies, do not put the
product to use and do not carry out any repairs
on your own.
Contact an authorized dealership or 

REBOTEC.

14.2 Maintenance electronic components

 

Regular safety checks are required: 

Check the wiring

Check 

the

charge 

level 

before 

each 

commissioning

Regularly  check  the  ventilation  openings  in  the
batteries.

Charge the battery modules regularly.

The batteries have a lifetime of 4 years depending
on the intensity of use and the general handling.

In the case of inconsistencies, do not operate the
unit and do not perform repairs on your own.
Please contact your dealer or 

REBOTEC.

Warning:  There  is  a  danger  of  injury  from 
slipping and pinching! 

14.3 Upkeep 

If  you  have  questions  concerning  maintenance
and  repairs  to  the  chair,  then  contact  an
authorized  dealer  or  medical  equipment  supply
outlet.

In  the  event  of  damage  and/or  needed  repairs,
contact the authorized dealership without delay.

Never  make  any  unauthorized  modifications  or
conversions  to  the  chair  or  the  electric  system.
These  will  adversely  affect  the  safety  and
function of the chair.

When  making  repairs,  use  only  original  spare
parts  from 

REBOTEC

  and  LINAK.

  Observe  all

advisories concerning guarantee and warranty.

Once  a  complete  maintenance  or  repair  of  the
chair is done, carry out a complete cleaning for
assuring 

hygiene. 

Follow 

the 

cleaning

instructions.

Follow the instructions for spare parts ordering,
re-use, and lifetime.

For documenting maintenance and repairs, 
REBOTEC offers a maintenance protocol. 
You may order User Instructions and 
maintenance protocol forms, or view them at 

www.rebotec.

com.au

Содержание Phoenix PS-E

Страница 1: ...subsequent sales The user instructions and the maintenance manual are downloadable on our homepage Electric Tilt Lift Shower Chair Phoenix Multi Seat with hygiene opening 352 xx 10 Serial Number Phoe...

Страница 2: ...4 4 Re use product lifetime 4 5 Disposal 4 6 Safety advisories for use 4 7 Product specifications 5 7 1 Delivery scope 5 7 2 Technical specifications 5 8 Spare parts accessories 6 9 Setting up the cha...

Страница 3: ...torage The product s temperature range for use must be between 41 F and 104 F Store the product in a dry room with a relative humidity of 30 50 The product may be used outside of these ranges but only...

Страница 4: ...nd its general condition 5 Disposal Disposal and recycling of used products as well as the packaging material must be done according to the applicable rules and legal regulations Please contact your l...

Страница 5: ...rst cool off before use When the chair is to be used by patients with insensitive skin persons who cannot feel temperature rises and or have skin damage an attendant must first check the chair for hea...

Страница 6: ...ional tolerances 5 mm 0 2 Schematic 8 Spare parts accessories Article No name 440 10 62 PUR seat pad 440 10 04 PUR care seat with care opening 309 10 Pan holder black 440 10 24 PUR care seat w o care...

Страница 7: ...A upwards and insert the tubes of the headrest unit into the openings Push the tubes down until the push buttons C click in the holes D Next push the button C all the way in and push the tube to the e...

Страница 8: ...delivered To remove it fold away the armrests as described in Section 11 8 on page 12 and pull the loops down from the armrest locking mechanism When re installing be sure the safety belt is not twis...

Страница 9: ...pply LED 1 ON green If the LED green lights up the wall charger is ready for operation LED 2 CHARGE not illuminated No battery in the charger or the battery is fully charged LED 2 CHARGE orange If the...

Страница 10: ...on class IPX5 IPX6 all electrical components must be installed with each other and none of the connections C must be open When connecting the manual switch make sure that the female connector D groove...

Страница 11: ...ed up 1 The patient must be positioned as near to the chair as possible By firmly holding on to the armrests the patient must then slowly sit down onto the seat If this is not possible due to physical...

Страница 12: ...ur foot 1 The wheel is now braked and the castor locked against swiveling 2 The wheel and castor are released by pressing the lever upwards with your foot 2 11 6 Removing the seat pad The seat with I...

Страница 13: ...Castor brakes do not work Generally defective See maintenance Castor wheels turn with difficulty Caked gummed up Clean Castor wheels do not turn Generally defective See maintenance Wheels wobble Uneve...

Страница 14: ...bility check function of the wheels replace defective wheels Function of the castor locking brakes replace defective brakes Function of the headrest Function of the chair inclination Check accessories...

Страница 15: ...REBOTEC BA No 02 07 147 2017 01 GB 15...

Страница 16: ...o 02 07 147 2017 01 GB 16 Notes We also supply Walking aids Shoulder crutches Shower stools Shower and commode chairs Multifunction Wheel Chairs Rollators Patient lifting hoists Standing up aids Your...