RBR#0008818revA
–
20
2.
a.
3.
4.
5.
6.
a.
7.
a.
b.
c.
d.
e.
f.
Tip the instrument onto its side and flush with ~2mL of oil.
The purpose of having the instrument on its side is to remove debris and contaminated oil.
Return the instrument to the vertical position, leaving the needle inserted into the buffer tube.
Continue to push more oil into the buffer tube, while slowly removing the needle.
Ensure you have filled the cavity with oil to the top of the threaded hole.
Clean the M6 set screw with 0.7mm
The 0.7mm hole is the smallest opening in the assembly, and if there is debris it may block this
area. You may need to clear this hole with a pick, or compressed air if available.
Reassemble, as per Figure 1 and instructions below:
Install the set screw with the 0.7mm hole (0004579) using the 3mm hex key. When installing the
set screw, the oil will be displaced and spill into the mating surface for the lower mesh insulator
disk.
Remove excess oil with a tissue or swab.
Once cleaned, install the Lower Mesh Insulator Disk (0004591) with the recess facing up.
The nickel-copper mesh (0004583) will rest in the recess of the lower mesh insulator disk, see
Figure 1.
The upper mesh insulator disk (00044592) is then placed on top of the lower mesh insulator disk
and nickel-copper mesh assembly.
Lastly, the retaining ring (0002069) is re-inserted by opening the split ring and feeding the ring
into the recession of the SEC, holding it in place with one finger while continuing to feed the
remainder of the ring into the recess.
Figure 1 - Anit biofouling mesh assembly
Содержание quartz3 Q
Страница 1: ...rbr global com INSTRUMENT GUIDE Q...
Страница 5: ...RBR 0008818revA 5 3 Revision history Revision No Release Date Notes A 04 May 2020 Original...
Страница 25: ...RBR 0008818revA 25 7 Appendix 7 1 Appendix A Product drawing...
Страница 26: ...RBR 0008818revA 26...
Страница 27: ...RBR 0008818revA 27 7 2 Appendix B Paros interface drawing...
Страница 28: ...RBR 0008818revA 28 7 3 Appendix C Anti biofouling mesh assembly...
Страница 29: ...RBR 0008818revA 29 7 4 Appendix D Calibration adapter installation...