6 Instrument Hardware
6.1 Opening and closing the logger
Instruments have both a "sensor end cap" and a "battery end cap". Only the battery end cap needs to be opened, the
data connector and battery door are inside. It may be unscrewed by hand or by using a rod inserted into the mooring
line hole for leverage. The end cap does not unscrew completely until the O-ring is clear of its mating surface in the
body. This ensures that it will not eject at high speed should the interior of the logger become pressurized.
Closing the logger is exactly the reverse of opening it. Remember to keep the O-rings clean and avoid scratching the O-
ring mating surfaces. Carefully inspect the O-rings, as described in the Inspecting the
the logger. Do not use excessive force when tightening the end cap; hand-tight is quite sufficient, as the seal depends
upon the O-rings, not the end cap tightness.