M
P
2
4
0
:
H
IG
H
I
S
O
L
A
T
IO
N
M
U
LT
IP
L
E
X
E
R
[ROUTe]:SELEct <int>
[ROUTe]:SELEct?
No response
<int>
Parameter <int>
The channel to operate on, 0 to 4
Description
The command exclusively sets a route
in both high and low relay banks. Any
other relays that were previously
enabled are disabled.
If called on channel 0, all relays are
disabled
The query form returns 0 if no relays
are enabled, 1 to 4 if a route is
exclusively set, -1 if more than one
route is set, and -2 if the high and low
relay banks are differently configured.
Example
SELE 1