Version 1.1 (released 14 April 2021)
http://razorbillinstruments.com
Page 22 of 40
M
P
2
4
0
: H
IG
H
IS
O
L
A
T
IO
N
M
U
LT
IP
L
E
X
E
R
Commands fall into two categories. Ones which end in a
question mark, and ones which do not. The former are referred to
as
queries
. Queries will always generate a response from the
instrument, and will never have any other effect. Commands
which are not queries will never generate a response from the
instrument, and generally do have an effect on it.
Normally, individual commands will be sent, each followed by a
terminator. However, it is also possible to chain commands with
semicolons, e.g.
H1 1;H2 1<\n>
in which case commands will be executed in order from left to
right. When chaining commands, care must be taken that the
complete chain, including terminator, is no longer than 255
characters. Otherwise the input buffer will overflow and some of
the commands will be lost.
A command which is not recognised by the instrument, is
followed by a parameter of the wrong type, or otherwise can’t be
understood by the instrument will generate an error. Errors are
stored in a queue, first in, first out, and as long as there is
something in the queue the error LED on the front panel will be
lit. The
SYSTem:ERRor?
command will fetch to error at the
top of the queue, and the
SYSTem:ERRor:COUNt?
command
will return the number of errors present. Note that in some cases,
one malformed command can generate several errors.
Command list
The commands listed here are supported by the 1.0.0 version.
Future firmware versions may add more commands and
functionality. An up-to-date manual will be made available on
the Razorbill Instruments website with each firmware update.