The correction formula:
Rs
=
Rm
Ts
+
Tk
Tm
+
Tk
Rs
Resistance corrected
Rm
Measured resistance
Ts
Desired reference temperature
Tm
The temperature at which resistance was/is measured
Tk
234.5 for copper
225.0 for aluminum
(aluminum may be as high as 240)
Alternative: Correction with “alpha25” to calculate Tk with the formula:
Tk
=
1
alpha25
−
25
When Temperature correction is activated the Main Test screen appears as below:
The “T” in the measuring channel box shows
that the displayed resistance value is corrected.
In this example it is corrected to 25 °C using Tk
of Copper and referring to Temperature probe 1.
Instruction Manual Winding Resistance Meter
Type WR100 / WR50 / WR14
Page 65 of 139