Wuhan Raycus Fiber Laser Technologies Co., Ltd
User Guide
RFL-C2000S-HP /RFL-C3000S-HP /RFL-C4000S-HP/RFL-C6000S-HP
26
EMON
Start Emission
– Turn on the laser (or enable it)
Send
:
“EMON\r”
Return
:
“EMON\r”
MPWROFF Main Power OFF
– Turn off the main power
Send
:
“MPWROFF\r”
Return
:
“MPWROFF\r”
MPWRON Main Power ON
– Turn on the main power
Send
:
“MPWRON\r”
Return
:
“MPWRON\r”
SPW
Set Pulse Width
Send
:
“SPW 100\r”
Return
:
“SPW
:
100\r”
(
Setpulsewidth100ms
)
Other return value
:
“ERR
:
input Err\r”
(
input pulse width<0.0001
)
“ERR
:
Out of Range\r”
(
Exceeds the maximum pulse
width range
)
“ERR
:
Duty Cycle too High\r”
(
The
set of duty cycle too high
)
“ERR
:
Duty Cycle too Low\r”
(
Set
pulse width <0.02
)
“SPW
:
100,Duty=100%\r”
(
Working in continuous mode
)
SPRR
Set Pulse Repetition Rate
Send
:
“SPRR 1000\r”
Return
:
“SPRR
:
1000\r”
Other return value
:
“ERR
:
input Err\r”
(
Input frequency <0
)
“ERR
:
Out of Range\r”
(
Exceeds the maximum pulse
width range
)
“ERR
:
Duty Cycle too High\r”
(
The
set of duty cycle too high
)
“ERR
:
Duty Cycle too Low\r”
(
Set
pulse width <0.02
)
“SPW
:
100,Duty=100%\r”
(
Working in continuous mode
)
SDC
Set Diode Current
– Set the operating current of the optical
module (unit:%).The setting value must
be lower than 100% and higher than the
minimum current setting value, which
can be set to 0. If the set value is bigger
than 100, the default is input 100.
Send
:
“SDC 100\r”
Return
:
“SDC
:
100\r”
Otherreturn value
:
“ERR
:
Input Err\r”
(
input value <0
)
“Laser is worked in AD Mode\r”
(
Working in external AD mode,
the command is invalid
)
RCS
Read Current Setpoint
Send
:
“RCS\r”
Return
:
“RCS
:
56.7\r”
(Present current setting value is
56.7%)