ERVICE
S
26
6823346-R3
FK0254.eps
3.1-4
View Current Calibration
Settings
NOTE:
Some software versions may not offer this
parameter.
1
Access the Parameters Menu as described in Section
3.1-1.
6
Each corner calibration value will show for two seconds.
3
Change the current parameter menu number to "92".
Refer to "Changing A Parameter's Settomg", Section
3.1-2, Steps 2 and 3.
2
Verify that Parameters 1 through 5 are correct according
to US (EU) values found on the sticker on the display.
Refer to "Changing A Parameter's Setting", Section
3.1-2, Steps 2, 3 and 4. Press the "Zero" button after
viewing each value.
4
The display will show "FC_bE".
5
Press the "Tare" button (down) to view the four corner
calibration values.
Corner A value:
Corner B value:
Corner C value:
Corner D value:
Procedure continued on next page
7
Press the "Total" button (up) to view the zero calibration
and weight calibration values.
1
3
4
5
7
6