NINOX 640 SU/USER MANUAL/01-21/REV1.0
point has been configured, selecting start will initiate the coolant re-circulation. For
more detailed information on all the operating modes of the
T257P Precision Chiller,
please refer to the Thermotek user manual.
WARNING: Please ensure the temperature set point of the chiller is above your ambient
dew point, otherwise condensation can form around the sensor package and cause
damage.
6.5 Draining the Chiller, Camera and Tubing
If using the Thermotek T257P Precision Chiller, it is recommended
that a coolant
change be implemented at a minimum, every six months, to keep the system in working
order. Based on the amount of use, you may need to perform this more frequently. For more
detailed information, please refer to the Thermotek user manual. If using another model of
chiller, please refer to its user manual for the recommended time frame for coolant changes.
If finishing the testing of a demo kit, please disconnect the tubing and drain all of the camera,
chiller and tubing before sending the equipment back to Raptor. Shorter draining tubes will
be supplied with the demo kit. Two draining tubes (with the appropriate mating connectors)
need to be connected to both connecters on each of the camera, chiller and tubing to drain
each individually.