4 MODIFY THE CORRECTION FACTOR IN FIELD
PLEASE CAREFULLY FOLLOW THE PROCEDURE INDICATED BELOW.
METER SETTING
2
3
4
5
Wait for the meter to go to standby.
Reset the resettable total.
Start dispensing into a measuring glass.
Stop dispensing when over 5 Litres of volume is reached, read
out the actual value. The volume that is displayed on the LCD
is the Display Value, not the Actual Value which may be slightly
higher. For example, in the figure on the right, the Display
Press the MENU key. Keep it pressed until showed as the right fig.,
the digit flash in æ zone, Press the RESET key to choose the right
digit from 0 to 9. Press the MENU key to go the the next digit so
Make sure the correction factor is right and then press the MENU
key. Keep it pressed until calibration is finished and the factor is
save. The meter will then return to standby.
Value is 18.96 while the Actual Value is 20.75.
1
that the Actual Value can be input.
7