4 MODIFY THE CORRECTION FACTOR IN FIELD
PLEASE CAREFULLY FOLLOW THE PROCEDURE INDICATED BELOW.

METER SETTING

2

3

4

5

Wait for the meter to go to standby.

Reset the resettable total.

Start dispensing into a measuring glass. 

Stop dispensing when over 5 Litres of volume is reached, read 

out the actual value.  The volume that is displayed on the LCD 

is the Display Value, not the Actual Value  which may be slightly

 higher. For example, in the figure on the right, the Display

Press the MENU key. Keep it pressed until showed as the right fig., 

the digit flash  in æ zone, Press the RESET key to choose the right 

digit from 0 to 9. Press the  MENU key to go the the next digit so 

Make sure the correction factor is right and then press the MENU 

key. Keep it pressed until calibration is finished and the factor is 

save. The meter will then  return to standby.

Value is 18.96 while the Actual Value is 20.75.

1

that the Actual Value can be input.

7

+49 (6158) 92980-0

Содержание DZP 20-Z

Страница 1: ...so that they can be read when required Translation of the original operation manual GB Stand 06 2018 i Weiterentwicklung im Sinne des Fortschrittes behalten wir uns vor Technical Modifications are su...

Страница 2: ...oved Always wear safety cloths respecting the applicable regulations for accident prevention Only use this device conforming to its purpose and function Improper use can cause severe injuries Caution...

Страница 3: ...meter features two buttons MENU and RESET wich individually perfprm two main functions and together other secondary functions The main functions perfomed are For the reset key resettung the partial r...

Страница 4: ...re parts and accessories 3 4 5 6 2 1 7 8 9 No Description Qty 1 Nozzle cover 1 2 Screw 4 3 Meter cover 1 4 LCD Display 1 5 Battery AAA 2 6 Meter base 1 7 Nozzle body 1 8 Screw 1 9 Spout 1 4 49 6158 92...

Страница 5: ...s MENU and RESET together ands hold for two seconds Value 1 4000 is the vorrection factor which can be reset 1234567 is the general total which can not be reset Measurement unit change see fig 4 Press...

Страница 6: ...ctor is right press the MENU key Keep it pressed until quit calibration mode the factor is saved correction factor modification mode factor can be changed one by one The meter goes to standby again ME...

Страница 7: ...he Actual Value which may be slightly higher For example in the figure on the right the Display Press the MENU key Keep it pressed until showed as the right fig the digit flash in zone Press the RESET...

Страница 8: ...0 1 In case of insufficient maintenance faults on operation use of not adequate spare parts or attachments all liabilities and rights of claim under guarantee are void 10 2 The manufacturer is not lia...

Отзывы: