Wait for the meter to go to standby.
Reset the resettable total.
Press the MENU key. Keep it pressed until similar with the image
showed (the digit flash in
①
zone), it means the meter is in
Press the RESET key to choose the right digit from 0 to 9. Press
the MENU key to start the next digit. So the digit of correction
Make sure the correction factor is right, press the MENU key.
Keep it pressed until quit calibration mode, the factor is saved.
correction factor modification mode.
factor can be changed one by one.
The meter goes to standby again.
METER SETTING
3 PROCEDURE FOR ENTER THE CORRECTION FACTOR DIRECTLY
Carefully follow the procedure indicated below.
Actual value 20.75
Display value 18.96
Current correction factor 1.000
Proper correction factor = current correction factor×(actual value/ display value)
FORMULA
Proper correction factor 1.000×(20.75/18.96)=1.000×1.094=1.094
6