INTRODUCTION, OPERATION & MAINTENANCE

Meter Scale Ranges:*

  

kV:  0 to ±199.9 kVDC

 

m

A:  0 to ±1999 

m

A

 k

W

:  1 to 1999 k

W

 (.001 to 1.999 M

W

 M

W:

 00.1M

to 199.9 M

 

 G

W

: .1G

W

 to 19.99 G

 

Note:

 For readings below 2 M

W

, more resolution 

can be obtained by using the k

W

 scale.

MULTI-FUNCTION 

ELECTROSTATIC METER

The Ransburg 

Multi-Function Electrostatic 

Spray Meter

 uses one meter to serve multi-func-

tions in electrostatic spray finishing. The meter 

can be used to measure conductivity, paint resis-

tance, short circuit current, resistance, and high 

KITS AND ADAPTERS

Part #

Description 

76652-01 

Kit for measuring high voltage. 

 

 

Includes 76634-00 Multi-Function Meter and 76667 High Voltage Probe Assembly.

76652-02 

Kit for measuring short circuit current (SCI),resistance and spray-ability.

  

 

Includes 76634-00 Multi-Function Meter and 76664-00 Test Lead Assembly.

76652-03 

Kit for measuring paint resistivity.  

 

Includes 76634-00 Multi-Function Meter and 7922-00 Paint Probe Assembly.

76652-04 

Deluxe Kit Performs all functions listed above. 

 

 

Includes 76634-00 Multi-Function Meter, 7922-00 Paint Probe Assem-bly, 76664-00 Test 

Lead Assembly, and 76667-00 High Voltage Probe Assembly.

A11759-00 

High Voltage Probe Adapter

 

Adapter to measure the voltage coming from an individual probe (electrode) assembly 

on all RMA-MMA Indirect Charge Robot Mounted Rotary Applicator,  Aerobell 2.5 Indirect 

Charge (A10924), and Aerobell II Indirect Charge (77603). 

voltage. The meter is available in individual kits 

focused on each function or a deluxe kit, which 

includes accessories to perform all the functions 

listed above.

 

The Ransburg Multi-Function Electrostatic Spray 

Meter is powered by a single 9V alkaline battery.  

The solid state circuitry only requires 9V input 

to power the meter display and condition all the 

signals from the accessory items.

Test Equipment - Introduction, Operation & Maintenance

TE-98-01.7

6

Содержание 76652-01

Страница 1: ...MODEL 76652 01 HIGH VOLTAGE PROBE 76652 02 SPRAYABILITY SCI METER 76652 03 PAINT RESISTIVITY METER 76652 04 DELUXE KIT A11759 00 HIGH VOLTAGE PROBE ADAPTER IMPORTANT Before using this equipment carefu...

Страница 2: ...manual has been changed from revision TE 98 01 6 to revision TE 98 01 7 Reasons for this change are noted under Manual Change Summary on page 23 of this manual Paint High Voltage SCI Test Equipment T...

Страница 3: ...RAYABILITY SCI AND RESISTANCE MEASUREMENT KIT 8 12 PAINT RESISTANCE MEASUREMENT KIT 13 14 HIGH VOLTAGE MEASUREMENT KIT 14 16 GENERAL MAINTENANCE 17 WARRANTY 18 WARRANTY POLICY 18 PAINT AND SOLVENT SPE...

Страница 4: ...inevitable Compare this manual with your system installation draw ings and appropriate Ransburg equipment man uals to reconcile such differences Carefulstudyandcontinueduseofthismanualwill provide a b...

Страница 5: ...mum of 5 C 9 F greater than the ambient temperature It is the end users responsibility to insure this condition is met Spray booth ventilation must be kept at the rates required by NFPA 33 OSHA countr...

Страница 6: ...ammable or combus tible materials The current overload sensitivity if equipped MUST be set as described in the correspond ing section of the equipment manual Protec tion against inadvertent arcing tha...

Страница 7: ...nding straps on wrists or legs may be used to assure adequate ground contact Operators must not be wearing or carrying any ungrounded metal objects When using an electrostatic handgun operators must a...

Страница 8: ...n hazard ous locations the power supply control cabinet and all other electrical equipment must be locat ed outside Class I or II Division 1 and 2 hazard ous areas in accordance with NFPA 33 and EN 50...

Страница 9: ...76652 03 Kit for measuring paint resistivity Includes 76634 00 Multi Function Meter and 7922 00 Paint Probe Assembly 76652 04 Deluxe Kit Performs all functions listed above Includes 76634 00 Multi Fu...

Страница 10: ...tion Meter Kit Parts 20913 00 CONTACT ROD 3587 03 COMPRESSION NUT 76634 00 MULTI FUNCTION TESTER 76667 HIGH VOLTAGE PROBE ASSEMBLY 7922 00 PAINT PROBE ASSEMBLY 76664 00 TEST LEAD ASSEMBLY Test Equipme...

Страница 11: ...he two 2 probes on the meter to the stripped bare ends of a single 12 long wire see Figure 2 If desired the continuity of the wire may be verified by using a Volt Ohm meter If the meter reads zero the...

Страница 12: ...10 Suitable 760 MW or 7 GW 105 Suitable 1020 MW or 1 0 GW 100 Suitable 1 1 GW 98 Test Sample Part 1 3 GW 95 Test Sample Part 1 5 GW 90 Test Sample Part 2 0 GW to Infinity 85 or less Requires Surface P...

Страница 13: ...dial to the MW scale Read the resistance If the resistance is greater than 1 0 MW the part is notadequatelygrounded referenceNFPA 77 NFPA 33 3 For resistance checks set the meter to MW scale see Figu...

Страница 14: ...he meter Ensure the meter reads zero If it does not read zero consult the factory for repair of the meter 2 Place appropriate test resistor in the high voltage cable socket of the power supply See app...

Страница 15: ...metallic ends of the two 2 probes together If the me ter reads zero the meter is in calibration If it does not read zero verify the continuity of the probe assembly and re perform the check If the met...

Страница 16: ...tal particles therefore their presence may go unrecog nized by the meter When such formulations are placed in electrostatic ap plicators and high voltage applied the metal particles may become aligned...

Страница 17: ...easingdistance to ground such as adding coiled fluid tubes C A U T I O N HIGH VOLTAGE MEASUREMENT KIT High Voltage Measurement The Ransburg High Voltage Measurement Kit using the multi function meter...

Страница 18: ...0 OSHA prior to operating this equipment 1 Set the meter dial to the kV scale see Figure 7 Plug the high voltage probe into the side of the meter Figure 7 Measuring High Voltage On Electrostatic AIr S...

Страница 19: ...n 18 Read the voltage displayed on the meter 30 seconds or less 19 Turn the power supply off 20 Disengage the probe cable from the power supply 21 Restoretheapplicatorconnectiontothepower supply and r...

Страница 20: ...ticker noting the next date when calibration expires Removal of this seal will void calibration of the unit A certificate of calibration will also be shipped with each calibrated unit NOTE GENERAL MAI...

Страница 21: ...RE IN PLASTIC SHRINK WRAP OR ANY OTHER NON APPROVEDCOVERING WILLVOIDTHIS WARRANTY FLUID HANDLING One 1 year from date of purchase i e Totalizer CCV Valves etc AIR BEARING ROTATORS Fifteen thousand 15...

Страница 22: ...er part number on the Front Cover 2 Corrected A11759 00 High Voltage Probe Adapter part number in the Multi Function Electrostatic Meter section 3 Updated to new format products and pictures 4 Deleted...

Страница 23: ...pare Parts Manufacturing 1910 North Wayne Street Angola Indiana 46703 9100 Telephone 260 665 8800 Fax 260 665 8516 Technical Service Assistance 320 Phillips Ave Toledo Ohio 43612 1493 Telephone toll f...

Отзывы: