background image

Short Circuit Current - SCI 

(Guns)

The Ransburg Multi-Function Meter may be 

used to measure the short circuit current of an 

electrostatic applicator. The short circuit current 

measurement is important in troubleshooting 

problems with electrostatic power supplies and 

applicators. These tests should be performed 

when a noticeable decrease in transfer efficiency 

occurs. The measured short circuit current value 

should be compared to the value listed in the 

appropriate power supply or applicator manual. 

This comparison indicates if there are problems 

with the power source or the applicator.

Operation

1.  Set the meter dial to the µA scale (see Fig-

ure 5).  Plug the probe with two (2) alligator 

clamp leads into the side of the meter. Check 

the meter calibration by touching the metallic 

ends of the two (2) probes together. If the me-

ter reads zero, the meter is in calibration. If it 

does not read zero, verify the continuity of the 

probe assembly and re-perform the check. If 

the meter still does not read zero, consult the 

factory for repair of the meter.

 

†

Ensure that meter's scale is proper-

ly set  for the function in which it is being 

used.  Damage to equipment may result if 

improperly set.

 

†

Prior to starting the SCI tests, ensure the 

voltage at the power supply is turned off.

W A R N I N G

!

C A U T I O N

!

2.  Attach one of the alligator clamps to the elec-

trode (brush on the #2 process bell). Be sure 

the clamp makes good contact with the elec-

trode (small wires on the #2 brush).  Attach the 

second alligator clamp to ground, preferably 

the handle or bracket. 

3.  Apply voltage to the gun from the power supply.  

The meter should read between the ranges 

indicated in the power supply and applicator 

manuals.

Test Equipment - Introduction, Operation & Maintenance

TE-98-01.7

12

Содержание 76652-01

Страница 1: ...MODEL 76652 01 HIGH VOLTAGE PROBE 76652 02 SPRAYABILITY SCI METER 76652 03 PAINT RESISTIVITY METER 76652 04 DELUXE KIT A11759 00 HIGH VOLTAGE PROBE ADAPTER IMPORTANT Before using this equipment carefu...

Страница 2: ...manual has been changed from revision TE 98 01 6 to revision TE 98 01 7 Reasons for this change are noted under Manual Change Summary on page 23 of this manual Paint High Voltage SCI Test Equipment T...

Страница 3: ...RAYABILITY SCI AND RESISTANCE MEASUREMENT KIT 8 12 PAINT RESISTANCE MEASUREMENT KIT 13 14 HIGH VOLTAGE MEASUREMENT KIT 14 16 GENERAL MAINTENANCE 17 WARRANTY 18 WARRANTY POLICY 18 PAINT AND SOLVENT SPE...

Страница 4: ...inevitable Compare this manual with your system installation draw ings and appropriate Ransburg equipment man uals to reconcile such differences Carefulstudyandcontinueduseofthismanualwill provide a b...

Страница 5: ...mum of 5 C 9 F greater than the ambient temperature It is the end users responsibility to insure this condition is met Spray booth ventilation must be kept at the rates required by NFPA 33 OSHA countr...

Страница 6: ...ammable or combus tible materials The current overload sensitivity if equipped MUST be set as described in the correspond ing section of the equipment manual Protec tion against inadvertent arcing tha...

Страница 7: ...nding straps on wrists or legs may be used to assure adequate ground contact Operators must not be wearing or carrying any ungrounded metal objects When using an electrostatic handgun operators must a...

Страница 8: ...n hazard ous locations the power supply control cabinet and all other electrical equipment must be locat ed outside Class I or II Division 1 and 2 hazard ous areas in accordance with NFPA 33 and EN 50...

Страница 9: ...76652 03 Kit for measuring paint resistivity Includes 76634 00 Multi Function Meter and 7922 00 Paint Probe Assembly 76652 04 Deluxe Kit Performs all functions listed above Includes 76634 00 Multi Fu...

Страница 10: ...tion Meter Kit Parts 20913 00 CONTACT ROD 3587 03 COMPRESSION NUT 76634 00 MULTI FUNCTION TESTER 76667 HIGH VOLTAGE PROBE ASSEMBLY 7922 00 PAINT PROBE ASSEMBLY 76664 00 TEST LEAD ASSEMBLY Test Equipme...

Страница 11: ...he two 2 probes on the meter to the stripped bare ends of a single 12 long wire see Figure 2 If desired the continuity of the wire may be verified by using a Volt Ohm meter If the meter reads zero the...

Страница 12: ...10 Suitable 760 MW or 7 GW 105 Suitable 1020 MW or 1 0 GW 100 Suitable 1 1 GW 98 Test Sample Part 1 3 GW 95 Test Sample Part 1 5 GW 90 Test Sample Part 2 0 GW to Infinity 85 or less Requires Surface P...

Страница 13: ...dial to the MW scale Read the resistance If the resistance is greater than 1 0 MW the part is notadequatelygrounded referenceNFPA 77 NFPA 33 3 For resistance checks set the meter to MW scale see Figu...

Страница 14: ...he meter Ensure the meter reads zero If it does not read zero consult the factory for repair of the meter 2 Place appropriate test resistor in the high voltage cable socket of the power supply See app...

Страница 15: ...metallic ends of the two 2 probes together If the me ter reads zero the meter is in calibration If it does not read zero verify the continuity of the probe assembly and re perform the check If the met...

Страница 16: ...tal particles therefore their presence may go unrecog nized by the meter When such formulations are placed in electrostatic ap plicators and high voltage applied the metal particles may become aligned...

Страница 17: ...easingdistance to ground such as adding coiled fluid tubes C A U T I O N HIGH VOLTAGE MEASUREMENT KIT High Voltage Measurement The Ransburg High Voltage Measurement Kit using the multi function meter...

Страница 18: ...0 OSHA prior to operating this equipment 1 Set the meter dial to the kV scale see Figure 7 Plug the high voltage probe into the side of the meter Figure 7 Measuring High Voltage On Electrostatic AIr S...

Страница 19: ...n 18 Read the voltage displayed on the meter 30 seconds or less 19 Turn the power supply off 20 Disengage the probe cable from the power supply 21 Restoretheapplicatorconnectiontothepower supply and r...

Страница 20: ...ticker noting the next date when calibration expires Removal of this seal will void calibration of the unit A certificate of calibration will also be shipped with each calibrated unit NOTE GENERAL MAI...

Страница 21: ...RE IN PLASTIC SHRINK WRAP OR ANY OTHER NON APPROVEDCOVERING WILLVOIDTHIS WARRANTY FLUID HANDLING One 1 year from date of purchase i e Totalizer CCV Valves etc AIR BEARING ROTATORS Fifteen thousand 15...

Страница 22: ...er part number on the Front Cover 2 Corrected A11759 00 High Voltage Probe Adapter part number in the Multi Function Electrostatic Meter section 3 Updated to new format products and pictures 4 Deleted...

Страница 23: ...pare Parts Manufacturing 1910 North Wayne Street Angola Indiana 46703 9100 Telephone 260 665 8800 Fax 260 665 8516 Technical Service Assistance 320 Phillips Ave Toledo Ohio 43612 1493 Telephone toll f...

Отзывы: