4. After the setting is completed, press the “C/T” key to exit the setting interface. Rotate the main handwheel
for sectioning or trimming as needed.
When feeding is exceeding the allowable range, the motor will automatically stop running and
give a beeping alarm. Under this situation, please turn the Small Handwheel in the opposite
direction to return.
Note: 1. The rotation of the handwheel should be even, and the rotation speed should be compatible
with the hardness of the specimen. The speed should be slower when sectioning harder
specimens.
2. To ensure satisfactory section quality, the recommended practice is to evenly rotate the
handwheel clockwise for a full circle.
5. After operation, the Main Handwheel should be returned to the starting position and locked. The push
button on the handwheel is to lock the Specimen Clamp in the upper position, and the lock lever at the
bottom of the instrument is to lock the Specimen Head at any position.
Note:
= lock ;
= unlock
6. Carefully remove the blade from the holder and place it in the blade box following the laboratory safety
regulations. Take the specimen out from the clamp. Remove the section debris from the instrument.
Remove the waste tray and clean up the waste debris.
7. The instrument will automatically enter the sleep standby state after it is not operated for 5-6 minutes,
as indicated by the sleep mode indicator light. The instrument returns to the working state by gently
moving the handwheel or pressing any button.
Note: To extend the life of the instrument, please turn off the power so that the instrument does not stay
in the standby mode for an excessively long period of time.