129
POWER SUPPLY TROUBLESHOOTING
Outer Power Supply Troubleshooting
1.
POSSIBLE CAUSE:
Computer does not turn on, indicator on the camera
beam isn’t bright.
REASON:
Outer power socket is not secure, or does not have power
connection.
Power button on the cabinet is either turned off or is not
receiving power.
Ensure that the electrical outlet is working by testing it with
another device.
2.
POSSIBLE CAUSE:
After equipment installation, there is smoking and
burning smell in the cabinet when power is connected.
REASON:
Connecting with industrial power (380V), will burn the power
supply. Computer also will be damaged under this situation.
SOLUTION:
Ensure the voltage selection switch is set to match the power at
your location, before connection. You will need to change the
power supply if the above situation happens.
REASON:
Connecting with industrial power (380V), will burn the power
supply. Computer also will be damaged under this situation.
SOLUTION:
Change the computer power.
3.
POSSIBLE CAUSE:
Computer works, but camera beam doesn’t have power. A demo version is played when running
the 3D software program.
CHECK:
The indicator of power supply is bright, if it is not bright, pull out the main power plug from HUB
communication board and plug back in, then check if the indicator is bright.
REASON:
If the indicator is still not bright, the power supply may not be working.
SOLUTION:
Change the power supply.
110 - 120
Содержание 3DP4100
Страница 22: ...22 㐀 䰀椀昀琀椀渀最 椀渀搀椀挀愀琀椀漀渀 ...
Страница 37: ...37 ...
Страница 38: ...38 ...
Страница 73: ...73 Choose rim diameter choose rim diameter from the list Rim diameter range list ...
Страница 77: ...77 Individual adjustment of rear wheel Toe Individual adjustment of rear wheel thrust angle ...
Страница 92: ...92 ...
Страница 135: ...135 ...