Basic Operation
R&S
®
ZVA80
27
Quick Start Guide 1312.6808.62 ─ 02
Fig. 3-2: Transmission measurement with an R&S ZVA80
3.7 Troubleshooting
The table below lists possible errors and remedies.
Error
Possible cause
Remedy
No output signal, LED next to the
mains switch on the rear of the con-
verter panel off.
Converter not power-sup-
plied
Check power supply and fuse (see
ter 2.5.7, "Switching on the External Test
Set"
Switchover between low frequency
and high frequency mode fails
Control connection not
established or Port 1 and
Port 2 connectors inter-
changed
Check connecting "H/L Switch" cable (see
chapter 2.5.4, "Connecting the Control
Cable"
Measurement shows noise only
No supply power at the
diplexer unit
chapter 2.3, "Diplexer Connectors"
3.8 Special Considerations for Type 03 and 04 Systems
The R&S ZVA50 network analyzer of an R&S ZVA80 type 03 or 04 supports measure-
ments with external test sets in analogy to the R&S ZVA80-BU; the calibration procedure
and the measurements are identical. Refer to the description in the previous sections.
Troubleshooting