Command reference
R&S
®
ZNL/ZNLE
929
User Manual 1178.5966.02 ─ 19
PORTs
The related test port numbers (comma-separated list of integers)
THRoughs
Measued Throughs: comma-separated list of directed port num-
ber pairs <pno1-pno2>
TSTamp
Timestamp in UTC
LTSTamp
Timestamp in local time
CKIT
For each port, the name of the used calibration kit is stored with
the calibration.
The query can return one of the following:
the cal kit name
(string)
if available and unique
'Multiple'
if cal kit names are available but not unique, i.e. if <PhyPt> was
not specified and multiple cal kits were used (multi-port cals with
different connector types, merged cals)
<Index>
Not used
Example:
SENSe:CORRection:DATA:PARameter?
Unrestricted query. Result looks like this:
100000,8500000000,201,-10,LIN
SENSe:CORRection:DATA:PARameter? STARt
...
SENSe:CORRection:DATA:PARameter? STYPe
Query settings one by one.
SENSe:CORRection:DATA:PARameter? RATTenuation
Query receiver attenuations. Result looks like this:
1,0.000000,2,0.000000
Usage:
Query only
Manual operation:
See
[SENSe<Ch>:]CORRection:DATA:PARameter:COUNt?
Gets the number of active system error corrections for channel <Ch>.
For the R&S
ZNL/ZNLE the result is 0 (uncalibrated) or 1 (calibrated). The properties of
the active system error correction can be queried using
.
Suffix:
<Ch>
.
Channel number
Usage:
Query only
VNA command reference