VNA GUI reference
R&S
®
ZNL/ZNLE
487
User Manual 1178.5966.02 ─ 19
Use the buttons representing the calibration standards to start the corresponding cali-
bration sweeps.
If "Show Cal Kit Label" is enabled on the
of the VNA Setup dialog, an
additional "Calibration Info" dialog is displayed. In this case, the cal sweep is started
from this dialog.
"Don't Show this Dialog Again" has the same effect as disabling "Show Cal Kit Label".
A green checkmark indicates that the calibration data of a standard has been acquired
successfully. A green checkmark after the port symbol indicates that the minimum
number of calibration measurements for the port has been performed.
Tip:
●
For a TRL calibration, at least one line standard must be measured between any
pair of calibrated ports. See
Chapter 9.5.1.5, "TRL calibration"
●
If the selected calibration kit comprises a Sliding Match, then for every required
Match measurement either the Match or at least three positions of the Sliding
Match must be measured. See
Chapter 9.5.2.3, "Sliding Match standards"
Remote command:
[SENSe<Ch>:]CORRection:COLLect[:ACQuire]:SELected
See also:
[SENSe<Ch>:]CORRection:COLLect:LOAD:SELected
Restart Sweep on Std. Meas.
If this function is active, a new standard measurement initiates a new sweep, starting
at the beginning ("Start") of the sweep range: The sweep points for the calibration
sweep are in ascending order, like for an ordinary measurement.
If "Restart Sweep on Std. Meas." is inactive, the new standard measurement is started
at the current sweep point; the current sweep is continued as a calibration sweep.
Apply
Is enabled when sufficient data have been acquired for the calibrated ports and stand-
ards. The button starts the calculation of the system error correction and closes the
calibration wizard. The current instrument settings are stored with the correction data.
To avoid incompatibilities, older calibration data is deleted unless it has been transfer-
red into the calibration "Pool" using the "Calibration Manager" (see
Note: Checks during the calculation of correction data
Cal softtool