Command reference
R&S
®
ZNL/ZNLE
882
User Manual 1178.5966.02 ─ 19
REDuce
Cumulative moving averages of the real and imaginary parts of
each measurement result, provides the most effective noise sup-
pression for the "Real" and "Imag" formats and for complex trace
formats.
MOVing
Simple moving averages of the real and imaginary parts of each
measurement result; similar to REDuce, but with finite history.
Manual operation:
See
[SENSe:]BANDwidth...
The
[SENSe:]BANDwidth...
commands set the bandwidth of the IF filter (measure-
ment bandwidth).The forms
BANDwidth
and
BWIDth
are equivalent.
[SENSe<Ch>:]BANDwidth[:RESolution]
............................................................................882
[SENSe<Ch>:]BWIDth[:RESolution]
.................................................................................882
[SENSe<Ch>:]BANDwidth[:RESolution]:SELect
................................................................ 883
[SENSe<Ch>:]BWIDth[:RESolution]:SELect
......................................................................883
[SENSe<Ch>:]BANDwidth[:RESolution]
<ResBandw>
[SENSe<Ch>:]BWIDth[:RESolution]
<IF Bandwidth>
Defines the IF bandwidth of the analyzer (measurement bandwidth). Values between
1
Hz and 500
kHz can be set.
Suffix:
<Ch>
.
Channel number
Parameters:
<IF Bandwidth>
Bandwidths can be set in 1
–
1.5
–
2
–
3
–
5
–
7 steps. The ana-
lyzer rounds up any entered value between these steps and
rounds down values exceeding the maximum bandwidth.
Range:
See above
*RST:
10 kHz
Default unit: Hz
Example:
BAND 1.1
Set a IF bandwidth of approx. 1.1
Hz for channel 1.
BAND?
The analyzer returns the rounded bandwidth of 1.5
Hz.
Manual operation:
See
VNA command reference