Network and remote settings
R&S
®
ZNL/ZNLE
169
User Manual 1178.5966.02 ─ 19
► Press the "Device Indicator" button on the "Instrument Home Page" to activate or
deactivate the "LXI" status LED to the right of the button. A green "LXI" status LED
indicates that a LAN connection has been established; a red status LED indicates
an error, for example, that no LAN cable is connected. When a device is connect-
ing to the instrument, the "LXI" status LED blinks. The "Device Indicator" setting is
not password-protected.
The most important control elements in the navigation pane of the browser interface
are the following:
●
"LAN Configuration" opens the menu with configuration pages.
●
"Status" displays information about the "LXI" status of the instrument.
LAN configuration
The LAN configuration consists of three parts:
●
"IP configuration" provides all mandatory LAN parameters.
●
"Advanced LAN Configuration" provides further LAN settings.
●
"Ping Client" provides the ping utility to verify the connection between the instru-
ment and other devices.
IP configuration
The "LAN Configuration > IP configuration" web page displays all mandatory LAN
parameters and allows their modification.
The "TCP/IP Mode" configuration field controls how the IP address for the instrument
gets assigned (see also
Chapter 8.5.1.2, "How to assign the IP address"
For the manual configuration mode, the static IP address, subnet mask, and default
gateway are used to configure the LAN. The automatic configuration mode uses DHCP
server or Dynamic Link Local Addressing (Automatic IP) to obtain the instrument IP
address.
Changing the LAN configuration is password-protected. The default password is
Lxi-
WebIfc
(notice upper and lower case characters).
You can change the LAN password in the "N Remote" dialog box, see
Advanced LAN configuration
The "LAN Configuration > Advanced LAN Configuration" parameters are used as fol-
lows:
●
"mDNS and DNS-SD" are two additional protocols: Multicast DNS and DNS Ser-
vice Discovery. They are used for device communication in zero configuration net-
works working without DNS and DHCP
●
"ICMP Ping" must be enabled to use the ping utility.
●
"VXI-11" is the protocol that is used to detect the instrument in the LAN.
How to set up a network and remote control