GUI Reference
R&S
®
ZNB/ZNBT
719
User Manual 1173.9163.02 ─ 55
Remote control via GPIB
To control external devices via GPIB, you have to use an
Log Errors
Enables the transfer of error messages for external devices (e.g. connection errors) to
the error log. The error log appears in the "Info" dialog; see
Remote command:
n/a
6.17.4.2
External Power Meters Dialog
The "External Power Meters" dialog configures external power meters so that they can
be used for measurements and power calibrations.
Access:
SYSTEM – [SETUP] > "External Devices" > "Power Meters..."
Background information
Chapter 5.7.25, "External Power Meters"
The configuration of a new external power meter involves the following steps:
1. Connect the power meter to your R&S
ZNB/ZNBT using a LAN (VXI-11), GPIB, or
USB interface.
2. If the power meter is connected via LAN, enable
3. Select
and wait until the power meter appears in the table of
"Known Devices".
Setup Softtool