Direct Compensation
With "Direct Compensation", a frequency-dependent transmission factor is calculated;
see
"Auto Length and Loss vs. Direct Compensation"
Remote command:
[SENSe:]CORRection:COLLect:FIXTure:LMParameter[:STATe]
[SENSe<Ch>:]CORRection:OFFSet<PhyPt>:DFComp[:STATe]?
Prompt for Each Port
Determines how the R&S
ZNB/ZNBT performs the sweeps for a given termination type
●
If unchecked, it performs the sweeps for Open/Short without interruption, implicitly
assuming that
all
ports are terminated accordingly
●
If checked, it interrupts the data acquisition process after each port, which allows
you to modify the test setup (e.g. terminate the next measured port).
All Compensation On / All Compensation Off
Activates/deactivates the length and loss compensation for all ports.
Use the checkboxes in the "Active" column of the
to activate/deactivate
the length and loss compensation for selected ports.
Remote command:
[SENSe<Ch>:]CORRection:OFFSet<PhyPt>:COMPensation[:STATe]
Measurement Type
The "Open" , "Short", and "Open and Short" buttons bring up the "
" dialog that guides you through the actual fixture measurement. See
vs. Open and Short compensation"
Measure Fixture wizard
The "Measure Fixture" dialog guides you through the previously configured fixture
compensation measurements.
Figure 6-49: Measure Fixture dialog: Auto Length (and Loss)
To acquire the necessary data, proceed as indicated in the information area.
With Prompt for Each Port disabled, "Take" acquires data for all selected ports in one
go. Otherwise sweeps are taken port by port.
Offset Embed Softtool