Startup
R&S
®
SZU100A
32
Manual 1425.3061.02 ─ 04
The R&S
SMW adjusts the internal calibration routines to the R&S
SZU.
"Adjust All" is disabled, i.e. in this configuration, you cannot execute adjust-
ment routines for the R&S
SMW.
8. Select "Adjust External Frequency Converter A/B" for all R&S
SZU.
The R&S
SMW automatically calibrates the parameters of the externally con-
nected R&S
SZU.
9. When completed, check that no errors occurred.
If the internal adjustments detect any errors, see User Manual of R&S
SMW,
chapter Internal Adjustment Settings for further information on how to pro-
ceed.
7.5
Behavior of the R&S
SMW When Working with
the R&S
SZU
At startup, the R&S
SMW firmware takes control of the R&S
SZU almost entirely.
You can only select the analog I/Q connection (differential or single ended), provi-
ded you are working with only one R&S
SZU.
Behavior of the R&S
SMW When Working with the R&S
SZU