Instrument Function
R&S
®
SMJ100A
382
Operating Manual 1403.7458.32 ─ 14
Trigger Delay
Sets the trigger signal delay in samples on external triggering or on internal triggering
via the second path. This enables the R&S
SMJ to be synchronized with the device
under test or other external devices.
The parameter
displays the delay converted in time.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:TRIGger[:EXTernal<1|2>]:DELay
[:SOURce<hw>]:BB:ARBitrary:TRIGger[:EXTernal<1|2>]:TDELay
Actual Trigger Delay
Displays the time (in seconds) an external trigger event or a trigger event form the
other path is delayed by.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:TRIGger[:EXTernal<1|2>]:RDELay?
Trigger Inhibit
Sets the duration for inhibiting a new trigger event subsequent to triggering. The input
is to be expressed in samples.
In the "Retrigger" mode, every trigger signal causes signal generation to restart. This
restart is inhibited for the specified number of samples.
This parameter is only available on external triggering.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:TRIGger[:EXTernal<1|2>]:INHibit
Next Segment Trigger In
The section "Next Segment Trigger" In of the menu is only displayed if a multi segment
waveform is loaded.
The section comprises the settings necessary to adjust the trigger for the switch
between the segments of a multi segment waveform. The parameters displayed
depend on the selected trigger source.
Current Segment
Indication of the waveform segment (segment file name and index) that is currently
output.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:WSEGment?
[:SOURce<hw>]:BB:ARBitrary:WSEGment:NAME?
Next Segment
Selects the waveform segment to be output next. It determines the start segment when
switching on the ARB and enables switchover to any other segment.
Baseband Signal - Baseband Block
Содержание SMJ100A
Страница 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Страница 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Страница 139: ...Manual Operation R S SMJ100A 123 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 140: ...Manual Operation R S SMJ100A 124 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 454: ...Instrument Function R S SMJ100A 438 Operating Manual 1403 7458 32 14 Baseband Signal Baseband Block...
Страница 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Страница 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Страница 843: ...Hardware Interfaces R S SMJ100A 827 Operating Manual 1403 7458 32 14 Figure A 2 BERT circuit diagram BERT Connector...
Страница 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...