Operating the instrument
R&S
®
RTE
101
Getting Started 1326.1026.02 ─ 10
Tap the icon and then tap the area or diagram to be deleted, or the waveform to
be switched off.
Save Settings
Saves the current instrument settings in a saveset. The filename is created
according to the autonaming pattern. You can reload the saveset using the "Load
saveset (Graphical recall)" toolbar icon, or using [FILE] > "Save/Recall" > "Set-
tings".
Save Screenshot
Saves a screenshot of the current display using the settings defined in
"File" menu > "Print Setup".
Create Report
Creates a report of the current measurement settings and results using the
settings defined in "File" menu > "Report Setup".
Clear all
Deletes all measurement results including long term measurement and
statistics, all waveforms, and the history.
Autoset and Preset
Performs an autoset, or a preset to a default state. The icons have
the same functionality as the corresponding keys on the front panel. They are
useful when you control the instrument remotely.
Run / Stop and Run Single
Starts and stops the continuous acquisition, or starts a defined
number of acquisition cycles. The icons have the same functionality as the corre-
sponding keys on the front panel. They are useful when you control the instru-
ment remotely.
Find Trigger Level
Analyses the signal and sets the trigger level to the middle of the signal
peaks.
Force Trigger
Starts an immediate single acquisition. If the acquisition is running in nor-
mal mode and no valid trigger occurs, use "Force Trigger" to confirm that a signal
is available. Then you can use the displayed waveform to determine how to trig-
ger on it.