Measurement Types and Result Displays
R&S
®
NRX
110
User Manual 1178.5566.02 ─ 07
Reflection
Opens a dialog to measure reflection parameters.
The ratio of forward and reverse power is a measure for the matching of the load that
can be expressed as standing wave ratio (SWR), return loss or reflection coefficient.
Off ← Reflection
Disabled.
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:FEED<Channel>
"POWer:OFF"
Reverse Power ← Reflection
Reverse power in W or dBm.
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:FEED<Channel>
"POWer:REVerse"
Standing Wave Ratio (SWR) ← Reflection
1 - Reflection coefficient
Standing wave ratio =
1 + Reflection coefficient
See also
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:FEED<Channel>
"POWer:SWRatio"
UNIT<Measurement>:POWer:REFLection
Return Loss ← Reflection
Reverse power
Return loss =
Forward power
10 x log
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:FEED<Channel>
"POWer:RLOSs"
UNIT<Measurement>:POWer:REFLection
Reflection Coefficient ← Reflection
Reflection coefficient =
Forward power
Reverse power
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:FEED<Channel>
"POWer:RCOefficient"
UNIT<Measurement>:POWer:REFLection