Remote Control Commands
R&S
®
NRX
344
User Manual 1178.5566.02 ─ 05
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Setting parameters:
<auto>
ONCE
Only valid parameter for this command.
0
Return value if no calibration is in progress.
Usage:
Setting only
CALibration<undef>:ALL:ZERO:AUTO
<auto>
Applies to all connected power sensors. See
Suffix:
<undef>
.
1 to n
No suffix required.
Setting parameters:
<auto>
Usage:
Setting only
CALibration<Sensor>:ZERO:FAST:AUTO
<auto>
Effective for trace measurements.
Performs fast zeroing. Since the commands are processed very quickly, they are not
overlapping.
The command is not fully R&S NRP2 compatible. If port
1 is assigned to measure-
ment
1 and so on, the command works. If port
1 is assigned to measurement
2, for
example, enable the R&S NRP2 emulation using
"NRP2"
.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Setting parameters:
<auto>
Usage:
Setting only
CALibration<undef>:ALL:ZERO:FAST:AUTO
<auto>
Applies to all connected power sensors. See
CALibration<Sensor>:ZERO:FAST:
Zeroing