Performing measurements
R&S
®
NRPM
75
User Manual 1425.8663.02 ─ 07
Averaging Mode
Indicates that the R&S
NRPM3N works in manual averaging mode and enables
you to set the average count, also considered as averaging factor.
See
Chapter 10.8.1, "Configuring averaging"
<Mode> ← Averaging Mode
Displays the averaging mode.
"manual"
Operates in manual mode only.
Enter the average count under
Remote command:
[SENSe<Sensor>:]AVERage[:STATe]
<Value> ← Averaging Mode
Sets the average count.
Remote command:
Filter Terminal Control
Defines the output mode for the measurement results, denoted as termination
control.
See also
Chapter 10.5, "Controlling the measurement"
"Repeating"
Provides the measurement result only after the entire measure-
ment has been completed.
The number of measurement cycle repetitions corresponds to
the set averaging factor, and thus directly relates to the mea-
surement duration: the higher the averaging factor, the longer
measurement time.
"Moving"
Provides intermediate values to facilitate early detection of
changes in the measured quantity. In the settled state, that
means when the number of measurements specified by the
averaging factor has been performed, the sensor module pro-
vides a moving average.
Remote command:
[SENSe<Sensor>:]AVERage:TCONtrol
Trigger settings
Provides the trigger parameters. You can define the conditions that have to be ful-
filled for a triggered measurement.
Measurement applications