Putting into operation
R&S
®
NGE100B
14
Getting Started 5601.1337.02 ─ 08
Provide adequate airflow
Do not block the air intake at the front and side of the instrument or the exhaust at
the rear. Install the instrument on a location that allows sufficient space for air cir-
culation at the air intake and exhaust. Recommended spacing to non-heat pro-
ducing surface is at least 2.5 inches (63.5 mm) from the ventilation holes.
Applicable local or national safety regulations and rules for the prevention of acci-
dents must be observed in all work performed.
The instrument is designed for use in the following sectors: Industrial, residential,
business and commercial areas and small businesses.
The instrument is designed for indoor use only. Before each measurement, you
need to verify at a known source if the instrument functions properly.
To disconnect from the mains, the low-heat device socket on the back panel
has to be unplugged.
See
for the general data on the instrument specification. For more infor-
mation, see the instrument product brochure (PN: 5214.8748.12).
Table 5-1: General data on instrument specification
Mains nominal voltage
AC
115 V / 230 V (
±
10 %) 50 Hz /
60 Hz
Power consumption
Maximum input power
180 W
Mains fuses
(fuse size: 5 mm x 20 mm)
115 V AC
IEC 60127-2/5 T 5 A 250 V
230 V AC
IEC 60127-2/5 T 2.5 A 250 V
Temperature
Operating temperature range
0
°
C to + 40
°
C
Storage temperature range
- 20
°
C to + 70
°
C
Humidity
Non-condensing
5 % to 80 %
Display
3.5
"
(QVGA)
Rack mount capability
½
19
"
R&S HZC95 option
Dimensions
W x H x D
222 mm x 88 mm x 280 mm
(8.74 in x 3.46 in x 11.02 in)
Weight
R&S NGE102B
4.9 kg (10.80 lb)
R&S NGE103B
5.0 kg (11.02 lb)
Intended operation