Remote commands to perform measurements in MSRA mode
R&S
®
FSWP-B1
95
User Manual 1177.5862.02 ─ 09
MANual
The user-defined RBW is used and the (FFT) window length
(and possibly the sample rate) are adapted accordingly. The
RBW is defined using the
command.
FFT
The RBW is determined by the FFT parameters.
*RST:
AUTO
Example:
IQ:BAND:MODE MAN
Switches to manual RBW mode.
IQ:BAND:RES 120000
Sets the RBW to 120 kHz.
Manual operation:
See
[SENSe:]IQ:BWIDth:RESolution
<Bandwidth>
This command defines the resolution bandwidth manually if
is set to
MAN
.
Defines the resolution bandwidth. The available RBW values depend on the sample
rate and record length.
Parameters:
<Bandwidth>
refer to data sheet
*RST:
RBW: AUTO mode is used
Default unit: HZ
Example:
IQ:BAND:MODE MAN
Switches to manual RBW mode.
IQ:BAND:RES 120000
Sets the RBW to 120 kHz.
Manual operation:
See
[SENSe:]IQ:FFT:ALGorithm
<Method>
Defines the FFT calculation method.
Parameters:
<Method>
SINGle
One FFT is calculated for the entire record length; if the FFT
length is larger than the record length (see
), zeros are appended to the
captured data.
Configuring MSRA measurements