Remote commands to perform measurements in MSRA mode
R&S
®
FSWP-B1
97
User Manual 1177.5862.02 ─ 09
Example:
IQ:FFT:WIND:OVER 0.5
Half of each window overlaps the previous window in FFT calcu-
lation.
Manual operation:
See
[SENSe:]IQ:FFT:WINDow:TYPE
<Function>
In the I/Q Analyzer you can select one of several FFT window types.
Parameters:
<Function>
BLACkharris
Blackman-Harris
FLATtop
Flattop
GAUSsian
Gauss
RECTangular
Rectangular
P5
5-Term
*RST:
FLAT
Example:
IQ:FFT:WIND:TYPE GAUS
Manual operation:
See
TRACe:IQ:BWIDth
This command defines or queries the bandwidth of the resampling filter.
The bandwidth of the resampling filter depends on the sample rate.
Parameters:
<Bandwidth>
Default unit: HZ
Manual operation:
See
TRACe:IQ:RLENgth
<NoOfSamples>
This command sets the record length for the acquired I/Q data.
Increasing the record length also increases the measurement time.
Note
: Alternatively, you can define the measurement time using the
SENS:SWE:TIME
command.
Parameters:
<NoOfSamples>
Number of samples to record.
*RST:
1001
Example:
TRAC:IQ:RLEN 256
Configuring MSRA measurements