DIAGnostic:SERVice:STESt:RESult?
This command queries the self-test results.
Return values:
<Results>
String of data containing the results.
The rows of the self-test result table are separated by commas.
Example:
DIAG:SERV:STES:RES?
would return, e.g.
"Total Selftest Status:
PASSED", "Date (dd/mm/yyyy): 09/07/2004 TIME:
16:24:54", "Runtime: 00:06", "...
Usage:
Query only
SOURce<si>:TEMPerature:FRONtend
This command queries the current frontend temperature of the R&S
FSW.
During self-alignment, the instrument's (frontend) temperature is also measured (as
soon as the instrument has warmed up completely). This temperature is used as a ref-
erence for a continuous temperature check during operation. If the current temperature
deviates from the stored self-alignment temperature by a certain degree, a warning is
displayed in the status bar indicating the resulting deviation in the measured power lev-
els. A status bit in the
STATUs:QUEStionable:TEMPerature
register indicates a
possible deviation.
Suffix:
<si>
.
irrelevant
Return values:
<Temperature>
Temperature in degrees Celsius.
Example:
SOUR:TEMP:FRON?
Queries the temperature of the frontend sensor.
14.10.3
Working with Transducers
The following commands configure and control transducer factors.
Useful commands for transducer management described elsewhere
●
MMEMory:SELect[:ITEM]:TRANsducer:ALL
Remote commands exclusive to transducer management
[SENSe:]CORRection:TRANsducer:ADJust:RLEVel[:STATe]
............................................ 1268
[SENSe:]CORRection:TRANsducer:CATalog?
.................................................................1268
[SENSe:]CORRection:TRANsducer:COMMent
................................................................1269
[SENSe:]CORRection:TRANsducer:DATA
...................................................................... 1269
[SENSe:]CORRection:TRANsducer:DELete
....................................................................1269
[SENSe:]CORRection:TRANsducer:SCALing
..................................................................1269
Configuring the R&S
FSW
Содержание FSW Series
Страница 1: ...R S FSW Signal and Spectrum Analyzer User Manual User Manual Version 43 1173941102 2...
Страница 16: ...Contents R S FSW 16 User Manual 1173 9411 02 43...
Страница 1388: ...Maintenance R S FSW 1388 User Manual 1173 9411 02 43 c If necessary remove any excess moisture with a dry soft cloth Cleaning...