Remote Commands
R&S
®
FSW
1192
User Manual 1173.9411.02 ─ 43
Suffix:
<n>
.
<m>
Parameters:
<Span>
Frequency. The maximum span depends on the marker position
and R&S
FSW model.
*RST:
5% of current span
Default unit: Hz
Manual operation:
See
CALCulate<n>:DELTamarker<m>:FUNCtion:BPOWer[:STATe]
<State>
This command turns delta markers for band power measurements on and off.
If neccessary, the command also turns on a reference marker.
Suffix:
<n>
.
<m>
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Manual operation:
See
"Band Power Measurement State"
See
"Switching All Band Power Measurements Off"
14.8.3.12
n dB Down Marker
The following commands control the n dB down markers.
CALCulate<n>:MARKer<m>:FUNCtion:NDBDown
...........................................................1192
CALCulate<n>:MARKer<m>:FUNCtion:NDBDown:FREQuency?
...................................... 1193
CALCulate<n>:MARKer<m>:FUNCtion:NDBDown:QFACtor?
........................................... 1194
CALCulate<n>:MARKer<m>:FUNCtion:NDBDown:RESult?
..............................................1194
CALCulate<n>:MARKer<m>:FUNCtion:NDBDown:STATe
................................................ 1194
CALCulate<n>:MARKer<m>:FUNCtion:NDBDown:TIME?
................................................ 1195
CALCulate<n>:MARKer<m>:FUNCtion:NDBDown
<Distance>
This command defines the distance of the n dB down markers to the reference marker.
Suffix:
<n>
.
<m>
Analyzing Measurements (Basics)