Common Analysis and Display Functions
R&S
®
FSW
545
User Manual 1173.9411.02 ─ 43
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:BPOWer[:STATe]
CALCulate<n>:DELTamarker<m>:FUNCtion:BPOWer[:STATe]
Span
Defines the span (band) around the marker for which the power is measured. The span
is indicated by lines in the diagram.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:BPOWer:SPAN
CALCulate<n>:DELTamarker<m>:FUNCtion:BPOWer:SPAN
Power Mode
Defines the mode of the power measurement result.
For analog demodulation measurements, the power mode is not editable for AM, FM,
or PM spectrum results. In this case, the marker function does not determine a power
value, but rather the deviation within the specified span.
"Power"
The result is an absolute power level.
The power unit depends on the
"Relative
Power"
This setting is only available for a delta band power marker.
The result is the difference between the absolute power in the band
around the delta marker and the absolute power for the reference
marker (see
on page 344). The powers are sub-
tracted logarithmically, so the result is a dB value.
[Relative band power (Delta2) in dB] = [absolute band power (Delta2)
in dBm] - [absolute (band) power of reference marker in dBm]
"Density"
The result is a power level in relation to the bandwidth, displayed in
dBm/Hz.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:BPOWer:MODE
CALCulate<n>:DELTamarker<m>:FUNCtion:BPOWer:MODE
Switching All Band Power Measurements Off
Deactivates band power measurement for all markers.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:BPOWer[:STATe]
CALCulate<n>:DELTamarker<m>:FUNCtion:BPOWer[:STATe]
9.3.4.7
Demodulating Marker Values and Providing Audio Output (Marker Demodulation)
Access
: "Overview" > "Analysis" > "Marker Functions" > "Select Marker Function" >
"Marker Demodulation" > "Marker Demod Config"
Or
: [MKR FUNC] > "Select Marker Function" > "Marker Demodulation" > "Marker
Demod Config"
The R&S
FSW provides demodulators for AM, FM and PM signals. The demodulation
marker function sends the demodulated data at the current marker frequency to the
Marker Usage
Содержание FSW Series
Страница 1: ...R S FSW Signal and Spectrum Analyzer User Manual User Manual Version 43 1173941102 2...
Страница 16: ...Contents R S FSW 16 User Manual 1173 9411 02 43...
Страница 1388: ...Maintenance R S FSW 1388 User Manual 1173 9411 02 43 c If necessary remove any excess moisture with a dry soft cloth Cleaning...