Remote Commands
R&S
®
FSW
1077
User Manual 1173.9411.02 ─ 43
[SENSe:]MIXer<x>:THReshold
<Value>
This command defines the maximum permissible level difference between test sweep
and reference sweep to be corrected during automatic comparison (see
Suffix:
<x>
.
irrelevant
Parameters:
<Value>
<numeric value>
Range:
0.1 dB to 100 dB
*RST:
10 dB
Example:
MIX:PORT 3
Manual operation:
See
Mixer Settings
The following commands are required to configure the band and specific mixer set-
tings.
[SENSe:]MIXer<x>:FREQuency:HANDover
.................................................................... 1077
[SENSe:]MIXer<x>:FREQuency:STARt
.......................................................................... 1078
[SENSe:]MIXer<x>:FREQuency:STOP
...........................................................................1078
[SENSe:]MIXer<x>:HARMonic:BAND:PRESet
................................................................ 1078
[SENSe:]MIXer<x>:HARMonic:BAND
.............................................................................1079
[SENSe:]MIXer<x>:HARMonic:HIGH:STATe
................................................................... 1079
[SENSe:]MIXer<x>:HARMonic:HIGH[:VALue]
................................................................. 1080
[SENSe:]MIXer<x>:HARMonic:TYPE
............................................................................. 1080
[SENSe:]MIXer<x>:HARMonic[:LOW]
............................................................................ 1080
.................................................................................... 1081
[SENSe:]MIXer<x>:LOSS:TABLe:HIGH
..........................................................................1081
[SENSe:]MIXer<x>:LOSS:TABLe[:LOW]
.........................................................................1081
................................................................................... 1081
............................................................................................1082
[SENSe:]MIXer<x>:RFOVerrange[:STATe]
...................................................................... 1082
[SENSe:]MIXer<x>:FREQuency:HANDover
<Frequency>
This command defines the frequency at which the mixer switches from one range to
the next (if two different ranges are selected). The handover frequency for each band
can be selected freely within the overlapping frequency range.
This command is only available if the external mixer is active (see
Suffix:
<x>
.
irrelevant
Parameters:
<Frequency>
numeric value
Setting Basic Measurement Parameters