Remote Commands
R&S
®
FSW
1080
User Manual 1173.9411.02 ─ 43
Manual operation:
See
[SENSe:]MIXer<x>:HARMonic:HIGH[:VALue]
<HarmOrder>
This command specifies the harmonic order to be used for the high (second) range.
Suffix:
<x>
.
irrelevant
Parameters:
<HarmOrder
numeric value
Range:
2 to 128 (USER band); for other bands: see band
definition
Example:
MIX:HARM:HIGH 2
Manual operation:
See
[SENSe:]MIXer<x>:HARMonic:TYPE
<OddEven>
This command specifies whether the harmonic order to be used should be odd, even,
or both.
Which harmonics are supported depends on the mixer type.
Suffix:
<x>
.
irrelevant
Parameters:
<OddEven>
ODD | EVEN | EODD
*RST:
EVEN
Example:
MIX:HARM:TYPE ODD
Manual operation:
See
[SENSe:]MIXer<x>:HARMonic[:LOW]
<HarmOrder>
This command specifies the harmonic order to be used for the low (first) range.
Suffix:
<x>
.
irrelevant
Parameters:
<HarmOrder>
numeric value
Range:
2 to 128 (USER band); for other bands: see band
definition
*RST:
2 (for band F)
Example:
MIX:HARM 3
Manual operation:
See
Setting Basic Measurement Parameters