Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
794
User Manual 1178.8520.02 ─ 01
14.5.6.3
Configuring a Multi-SEM Measurement
In the Spectrum application only, spectrum emissions can be measured for multiple
sub blocks of channels (see
Chapter 7.6.4.5, "SEM with Multiple Sub Blocks ("Multi-
on page 216). Up to three sub blocks (with two gaps) can be defined. For each
sub block, the familiar configuration settings concerning ranges, limit lines etc. can be
defined in individual tabs (select the sub block using the <sb> suffix in the correspond-
ing commands). In addition, settings on the sub blocks themselves must be configured.
Useful commands for multi-SEM measurements described elsewhere:
●
[SENSe:]ESPectrum<sb>:RANGe<ri>:MLCalc
Remote commands exclusive to multi-SEM measurements
..................................................................................794
...................................................................................794
[SENSe:]ESPectrum<sb>:SCENter
<Frequency>
This command defines the center frequency of the selected sub block in a Multi-SEM
measurement.
Suffix:
<sb>
.
1..n
Sub block in a Multi-SEM measurement
Parameters:
<Frequency>
Frequency within the currently defined global span (see
Range:
1 to 3
*RST:
1
Default unit: Hz
Example:
ESP1:SCEN 1GHZ
Manual operation:
See
[SENSe:]ESPectrum<sb>:SCOunt
<Subblocks>
This command defines the number of sub blocks in the SEM measurement.
Suffix:
<sb>
.
irrelevant
Parameters:
<Subblocks>
Number of sub blocks in the SEM measurement.
Range:
1 to 3
*RST:
1
Example:
ESP:SCO 2
Manual operation:
See
Configuring and Performing Measurements