Measurements and Results
R&S
®
FSVA3000/ R&S
®
FSV3000
228
User Manual 1178.8520.02 ─ 01
Sub Block / Center Freq
Defines the center frequency for an individual sub block. The center frequency deter-
mines the reference range used for each block.
For measurements with only one sub block, this setting corresponds to the global set-
ting in the "Frequency" settings (see
).
Remote command:
Standard / MSR Settings
Defines the use of a standard settings file or a multi-standard radio configuration for a
particular sub block. For details, see
Chapter 7.6.5.6, "Standard Files"
Chapter 7.6.5.5, "MSR Settings"
Note that
either
a standard
or
an MSR setting can be selected; if one is selected, the
other is disabled.
Remote command:
[SENSe:]ESPectrum<sb>:PRESet[:STANdard]
Chapter 14.5.6.3, "Configuring a Multi-SEM Measurement"
Edit Sweep List
Switches to the "Sweep List" tab of the "Spectrum Emission Mask" dialog box to con-
figure the individual frequency ranges and mask limits for the corresponding sub block.
See
Chapter 7.6.5.1, " Sweep List "
7.6.5.3
Reference Range
Access
: "Overview" > "Select Measurement" > "Spectrum Emission Mask" > "Refer-
ence Range"
The range around the center frequency is defined as the reference range for all other
ranges in the sweep list.
Spectrum Emission Mask (SEM) Measurement