Measurements on CW Signals
R&S FSH
1145.5973.12 2.14
E-14
Performing Distance-To-Fault Measurements
(Only for the R&S FSH with tracking generator, installed option R&S FSH-B1 (distance-to-fault
measurement) and VSWR Bridge and Power Divider R&S FSH-Z2 or R&S FSH-Z3).
R&S FSH with VSWR Bridge R&S FSH-Z3
Control
cable
RF input
VSWR Bridge
and
Power Divider
R&S FSH-Z2
Test port
Generator
output
Measurement
cable
Short
circuit
DUT
R&S FSH with VSWR Bridge
R&S FSH-Z2
Connect the control cable of the R&S FSH-Z2 or R&S FSH-Z3 to the power sensor connector of the
R&S FSH.
Connect the RF and generator port of the R&S FSH-Z2/R&S FSH-Z3 to the RF input and generator
output of the R&S FSH.
Connect the 1 m test cable supplied with option R&S FSH-B1 to the bridge test port.
Note:
The 1 m cable must be used. Results are invalid without this cable.
Press the MEAS key.
Press the MEASURE softkey.
Using the cursor keys or rotary knob, select DISTANCE TO FAULT from the menu and confirm with
the ENTER key or the MEAS softkey.
The R&S FSH switches on the distance-to-fault measurement function.
The R&S FSH delivers optimum results if the center frequency is set to the frequency at which the
device under test is operated.
Press the FREQ key.
Input the center frequency, e.g. frequency of the antenna at the end of the cable under test.
VSWR bridge
Measurement
cable
RF input
Control
cable
Generator
output
Short circuit
DUT
Test port