Specifications
¸FSH3
¸FSH6
Frequency
Frequency range
100 kHz to 3 GHz
100 kHz to 6 GHz
Reference frequency
Aging
1 ppm/year
Temperature drift
0 °C to +30 °C
+30 °C to +50 °C
2 ppm
in addition 2 ppm/10 °C
Frequency counter
Resolution
1 Hz
Counter accuracy
S/N > 25 dB
± (frequency × reference frequency error)
Frequency span
model .03/.23,
model .06/.26
model .13
0 Hz, 100 Hz to 3 GHz
0 Hz, 10 kHz to 3 GHz
0 Hz, 100 Hz to 6 GHz
–
Spectral purity
SSB phase noise
f = 500 MHz,
+20 °C to +30 °C
30 kHz from carrier
<85 dBc (1 Hz)
100 kHz from carrier
<100 dBc (1 Hz)
1 MHz from carrier
<120 dBc (1 Hz)
Sweep time
span = 0 Hz
1 ms to 100 s
span > 0 Hz
20 ms to 1000 s, min. 20 ms/600 MHz
Bandwidths
Resolution bandwidths (–3 dB)
model .13
1, 3, 10, 30, 100, 200, 300 kHz, 1 MHz
model .03/.23,
model .06/.26
in addition 100 Hz, 300 Hz
Tolerance
≤
300 kHz
±5 %, nominal
1 MHz
±10 %, nominal
Resolution bandwidths (–6 dB)
with option ¸FSH-K3
installed
in addition 200 Hz, 9 kHz, 120 kHz, 1 MHz
Video bandwidths
10 Hz to 1 MHz in 1, 3 steps
11
Handheld Spectrum Analyzer ¸FSH
Specifications apply under the following conditions: 15 minutes warm-up time at ambient temperature, speci-
fied environmental conditions met and calibration cycle adhered to. Data without tolerances: typical values.
Data designated as “nominal”: design parameters, i. e. not tested.