112
PD 1309.6269.62-02
Remove the SHORT from the measurement
cable.
Screw the cable under test to the measurement
cable.
The R&S FSH displays the return loss produced in
the cable under test versus the distance.
Checking the spectrum in the span for detecting
external interferers
Press the MEAS MODE softkey.
Select SPECTRUM using the rotary knob or the
cursor keys (
∧
or
∨
).
Confi rm your selection by pressing the MEAS
MODE softkey again or by pressing the ENTER
key.
The R&S FSH turns off the tracking generator and
displays the spectrum over the span of the DTF
measurement.
To indicate that the R&S FSH is in the spectrum
mode, DTF SPECTRUM is displayed in the upper
right-hand corner of the screen. Otherwise, the
R&S FSH uses exactly the same settings as it did
for DTF measurements.
GalaxyE_v1-10_3.indd 112
GalaxyE_v1-10_3.indd 112
08.04.2009 15:21:53
08.04.2009 15:21:53
Содержание FSH series
Страница 1: ...R S FSH Handheld Spectrum Analyzer Quick Start...
Страница 29: ...27 E 02...
Страница 72: ...70 PD 1309 6269 62 02...
Страница 130: ...128 PD 1309 6269 62 02...