![background image](http://html1.mh-extra.com/html/rands/fsh-series/fsh-series_quick-start-manual_3147226105.webp)
103
E-02
Terminate the measurement input of the bridge
or the end of the measurement cable into a
short circuit.
Start the short-circuit calibration by pressing
CONTINUE.
In the third step of calibration, terminate the
measurement port into a 50
Ω
termination.
Terminate the measurement input of the bridge
or the end of the measurement cable into a
50
Ω
termination (LOAD)
Start termination calibration by pressing
CONTINUE.
When calibration is done, the R&S FSH outputs
the message “Calibration Done” for three seconds.
REFL. VECTOR (CAL) is displayed in the upper area
of the diagram. The R&S FSH is vector-calibrated
for refl ection measurements.
Vector refl ection measurements
Select a measurement within the FORMAT softkey
menu. A detailed description of the measurements
is given in the Oerating Manual in chapter
“Measuring the Refl ection Magnitude”
.
GalaxyE_v1-10_3.indd 103
GalaxyE_v1-10_3.indd 103
08.04.2009 15:21:52
08.04.2009 15:21:52
Содержание FSH series
Страница 1: ...R S FSH Handheld Spectrum Analyzer Quick Start...
Страница 29: ...27 E 02...
Страница 72: ...70 PD 1309 6269 62 02...
Страница 130: ...128 PD 1309 6269 62 02...