Preparing for Use
R&S
®
ESW
20
Getting Started 1328.4922.02 ─ 06
To display the alignment results again later
●
Press the [SETUP] key.
●
Press the "Alignment" softkey.
Performing a selftest
The selftest does not need to be repeated every time the instrument is switched
on. It is only necessary when instrument malfunction is suspected.
1. Press the [SETUP] key.
2. Press the "Service" softkey.
3. Switch to the "Selftest" tab in the "Service" dialog box.
4. Select the "Start Selftest" button.
Once the instrument modules have been checked successfully, a message is
displayed.
4.1.7
Checking the Supplied Options
The instrument may be equipped with both hardware and firmware options. In
order to check whether the installed options correspond to the options indicated
on the delivery note, proceed as follows.
1. Press the [SETUP] key.
2. Press the "System Config" softkey.
3. Switch to the "Ve Options" tab in the "System Configuration" dialog
box.
A list with hardware and firmware information is displayed.
4. Check the availability of the hardware options as indicated in the delivery
note.
4.2
Windows Operating System
The instrument contains the Microsoft Windows operating system which has been
configured according to the instrument's features and needs. Changes in the sys-
Windows Operating System